File-Specific Cyclic Fatigue Resistance of NiTi Instruments After Repeated Use in Simulated Canals: Patterns Compatible with Potential Stress-Induced Martensite Transformation Effects

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q1
저널
MATERIALS19(5) · 2026
피인용
Web of Science 1Scopus 1
FWCI (Scopus)
0.85
JCR IF
JCR 2024 3.2
ISSN
1996-1944

저자

  1. 조현우제1저자Pusan Natl Univ, Dent & Life Sci Inst, Dept Conservat Dent, Sch Dent Dent,Res Inst, Yangsan 50612, South Korea
  2. 곽상원Pusan Natl Univ, Dent & Life Sci Inst, Dept Conservat Dent, Sch Dent Dent,Res Inst, Yangsan 50612, South Korea
  3. Ha, Jung-HongKyungpook Natl Univ, Sch Dent, Dept Conservat Dent, Daegu 41566, South Korea
  4. Sigurdsson, AsgeirNYU Coll Dent, Dept Endodont, New York, NY 10010 USA
  5. 김현철교신저자Pusan Natl Univ, Dent & Life Sci Inst, Dept Conservat Dent, Sch Dent Dent,Res Inst, Yangsan 50612, South Korea

초록

This study evaluated changes in the number of cycles to fracture (NCF) of Nickel–Titanium (NiTi) files after repeated use in simulated canals and investigated the potential relationship with stress-induced martensite (SIM) transformation effect. A total of 225 ProTaper Ultimate (PTUL) files were divided into three groups: Group 1 consisted of new files, Group 2 comprised files used to shape two resin simulated canals, and Group 3 consisted of files used to shape four canals. The simulated resin canals with a 16 mm length of J-shaped with 35° curvature were prepared using PTUL Slider, Shaper, F1, F2, and F3 files sequentially. After instrumentation, the cyclic fatigue resistance of each sequential file was assessed in a 35° curved steel canal by rotating at 400 rpm using a custom-made device. Statistical analysis was performed using one-way ANOVA with Tukey’s post hoc test or Kruskal–Wallis with Dunn’s test with Bonferroni correction for parametric and non-parametric data, respectively. Slider and Shaper maintained stable NCF across all groups (p > 0.05). In contrast, F1 showed a transient increase (117.7%) after two uses but declined significantly (91.6%) after four uses (p < 0.05). F2 and F3 demonstrated progressive NCF reductions (F2: 72.9%; F3: 71.5% after four uses), with F3 showing the most pronounced decline (p < 0.05). Repeated use of NiTi files reduced their cyclic fatigue resistance in a file-specific manner, with larger finishing files most affected. The distinctive F1 pattern suggests potential preload-related or SIM transformation effects that warrant further metallurgical investigation.

키워드

cyclic fatiguefile sequencesfracture risknumber of cycles to fracturereuseSIM transformation