Chronic In Vivo Biostability and Biocompatibility Evaluation of Polyether-Urethane-Based Balloon Implants for Cardiac Application in a Porcine Model

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q3
저널
BIOENGINEERING-BASEL13(2) · 2026
피인용
Web of Science 1Scopus 1
FWCI (Scopus)
1.12
JCR IF
JCR 2024 3.7
ISSN
2306-5354

저자

  1. 김민기제1저자Pusan Natl Univ, Dept Polymer Sci & Engn, Busan 46241, South Korea
  2. 서재영Pusan Natl Univ, Sch Med, Dept Internal Med, Busan 49241, South Korea
  3. 김준홍Pusan Natl Univ, Sch Med, Yangsan Hosp,Cardiovasc Ctr,Dept Internal Med & Ca, Res Inst Convergence Biomed Sci & Technol, Yangsan 50612, South Korea
  4. 김진창Pusan Natl Univ, Dept Clin Bioconvergence, Busan 46241, South Korea
  5. Park, Jun-YongTau Med Inc, Dept R&D Ctr, Busan 50612, South Korea
  6. Song, Hyun-ATau Med Inc, Dept R&D Ctr, Busan 50612, South Korea
  7. Song, Kyeong-DeokTau Med Inc, Dept R&D Ctr, Busan 50612, South Korea
  8. 천민구교신저자Pusan Natl Univ, Sch Med, Yangsan Hosp,Cardiovasc Ctr,Dept Internal Med & Ca, Res Inst Convergence Biomed Sci & Technol, Yangsan 50612, South Korea

초록

Polyurethane-based implantable devices (PUIDs) delivered via catheter are increasingly used in structural heart interventions; however, limited in vivo data exist regarding their long-term biostability and biological safety. This study evaluated a balloon-shaped implant made of Pellethane®, a polyether-based polyurethane, designed as a three-dimensional intracardiac spacer and deployed via percutaneous femoral vein access. The device was chronically positioned adjacent to the tricuspid valve annulus in seven pigs for 24 weeks. Explanted devices and surrounding tissues were evaluated through material characterization (SEM, GPC, FT-IR, and 1H-NMR) and histological analysis. SEM and FT-IR confirmed preserved surface morphology and chemical bonds, GPC showed stable molecular weight, and 1H-NMR revealed intact urethane and ether linkages. Materials characterization revealed no evidence of hydrolytic or oxidative degradation, indicating structural stability of the devices. Histological analysis showed stable device positioning with minimal thrombosis or inflammatory response. Biocompatibility was confirmed via ISO 10993-1:2018 Standard (International Organization for Standardization (ISO): Geneva, Switzerland, 2018), and extractable substances were evaluated under exhaustive extraction conditions specified by ISO 10993-18:2020 (International Organization for Standardization (ISO): Geneva, Switzerland, 2020), with no toxicologically significant findings. These findings support the long-term biostability and biological safety of the PUIDs in dynamic cardiac environments, informing future design criteria for catheter-delivered cardiovascular devices.

키워드

biocompatibilitybiostabilityPellethanepolyurethane