Multi-Method Assessment of Inter-rater Reliability in Dental Implant Stability Measurements

문서 유형
연구출판물 · 학술지 논문
등재
KCI
저널 순위
KCI Q3
저널
대한구강악안면임프란트학회지29(4) · pp.177 · 2025
피인용
KCI 0
ISSN
2765-7833

저자

  1. 구민주제1저자부산대학교
  2. 김석현부산대학교
  3. 양윤규부산대학교
  4. 오형석부산대학교
  5. 정해린부산대학교
  6. 허동우부산대학교
  7. 허중보교신저자부산대학교

초록

Purpose: Implant stability is a critical determinant of successful osseointegration. Common assessment methods include percussion testing, resonance frequency analysis, radiographic evaluation of marginal bone levels, and histometric analysis. Although many studies have assessed the reliability of individual methods, few have examined their concordance when applied by different examiners. This study aimed to evaluate the reproducibility of widely accepted implant stability measurement techniques by comparing results obtained by researchers who received identical training. Materials and Methods: Five junior researchers from a dental school assessed the stability of implants placed in one beagle. The methods included Implant Stability Quotient (ISQ), Implant Stability Value (ISV), radiographic analysis, and histologic analysis—specifically, bone-to-implant contact (BIC) and interthread bone density (ITBD). Interrater reliability was calculated using the intraclass correlation coefficient (ICC) with a 95% confidence interval. Results: All methods demonstrated statistically significant reliability (p < .05). BIC (ICC = 0.806) and ITBD (ICC = 0.854) exhibited high to very high reliability. ISQ (ICC = 0.774) and radiographic analysis (ICC = 0.638) showed moderate to high reliability. ISV (ICC = 0.447) exhibited low reliability. Conclusion: BIC, ITBD, ISQ, and radiographic analysis demonstrated consistent reproducibility among trained examiners, with BIC and ITBD showing particularly high concordance. ISV measurements were comparatively less reliable. Within the limitations of this study—including a small sample size, the use of a single animal model, and limited measurement repetitions—histometric parameters demonstrated the strongest interrater agreement

키워드

Dental implantImplant stabilityInterrater reliabilityResonance frequency analysis

식별자

KCI
ART003293904