Study to improve reproducibility and accuracy of DIBH for breast cancer radiotherapy: A laser-based self-monitoring system
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q4Scopus Q4
- 저널
- INTERNATIONAL JOURNAL OF RADIATION RESEARCH23(3) · pp.659 · 2025
- 피인용
- Web of Science 0Scopus 1
- FWCI (Scopus)
- 0
- JCR IF
- JCR 2024 0.4
- ISSN
- 2322-3243
저자
- 김동운제1저자Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan, South Korea
- 전호상Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan, South Korea
- 기용간교신저자Pusan Natl Univ, Yangsan Hosp, Dept Radiat Oncol & Res Inst, Yangsan, South Korea
- 주지현Pusan Natl Univ, Yangsan Hosp, Dept Radiat Oncol & Res Inst, Yangsan, South Korea
- 김원택Pusan Natl Univ, Sch Med, Yangsan, South Korea
- 김동현Pusan Natl Univ, Sch Med, Yangsan, South Korea
- 남지호Pusan Natl Univ, Sch Med, Yangsan, South Korea
- 박달Pusan Natl Univ, Sch Med, Yangsan, South Korea
초록
Background: Deep-inspiration breath-hold (DIBH) is a helpful technique during radiation therapy (RT) for breast cancer that can protect vital organs and deliver a more accurate treatment dose to the target. A laser sensorbased position monitoring system is developed to measure the absolute position of an object and achieve high accuracy and reproducibility during DIBH. Materials and Methods: A laser distance sensor (LDS) was fabricated to be mounted on a commercially available breast board and configured to provide real-time monitoring to assist with respiratory control. DIBH was measured in 10 volunteers with and without self-monitoring. Using an anthropomorphic phantom, we calculated the change in dose distribution due to DIBH error. We estimated the change in dose to target, the heart, and left lung due to DIBH error from the volunteer data. Results: With monitoring, the DIBH error was within 2 mm; without monitoring, the DIBH error increased to approximately 5 mm. Some of the volunteers who did not perform self-monitoring had large DIBH errors. This resulted in suboptimal dose distributions to the target, heart, and left lung, due to unintended alterations in the intended dose distribution. Conclusion: A self-monitoring system using LDS can greatly assist in the reproducibility of DIBH, thus helping to maintain the planned prescribed dose.
키워드
Breath-holdinglaser distance sensorbreast neoplasmscardiac sparingradiotherapyBREATHING ADAPTED RADIOTHERAPYRADIATION-THERAPYCLINICAL-EXPERIENCEHOLD DIBHHEARTDEVICE