Quantitative and Qualitative Evaluation of Radiation Dose and Image Quality at 100 kVp and 80 kVp in Lower Extremity CT Angiography
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCI
- 저널 순위
- KCI Q3
- 저널
- 한국산업융합학회논문집28(3) · pp.623 · 2025
- 피인용
- KCI 0
- ISSN
- 1226-833X
저자
초록
This study aimed to evaluate whether diagnostic image quality can be maintained while reducing radiation dose by lowering the tube voltage from 100 kVp to 80 kVp, and to assess the potential of this method as a standardized protocol. Sixty patients (30 per group at 100 kVp and 80 kVp) underwent lower extremity CTA using a multi-detector CT scanner (Somatom Definition Flash, Siemens, Germany). Objective image noise (standard deviation, SD) was measured in the abdominal aorta, inguinal region, and knee joint. Subjective image quality was rated by three radiologists on a 4-point scale. The average SD values were higher at 80 kVp than at 100 kVp (abdominal aorta: 35.84 vs. 27.48; inguinal region: 38.12 vs. 27.27; knee: 37.50 vs. 26.93). Despite the increased noise, subjective evaluation at 80 kVp yielded a mean score of 3.22, with 99.57% of images rated ≥2, indicating diagnostic adequacy. Additionally, a 54.5% reduction in radiation dose was achieved. Lower extremity CTA using 80 kVp produces diagnostically acceptable images with significantly reduced radiation exposure. This low-kVp protocol may be considered a feasible and universal strategy for long-range vascular imaging.
키워드
Lower extremity CTAMulti-detector CT scannerDiagnostic adequacy