Functional near-infrared spectroscopy sensor using a combined short-channel regression and frequency-division multiplexing
- 문서 유형
- 연구출판물 · 학술대회 발표논문
- 등재
- Scopus
- 저널 순위
- Scopus Q4
- 피인용
- Scopus 0
- FWCI (Scopus)
- 0
- ISSN
- 0277-786X
저자
- 박창현제1저자Department of Cogno-Mechatronics Engineering, Pusan National University, Busan, 46241, Korea
- 조동영Department of Cogno-Mechatronics Engineering, Pusan National University, Busan, 46241, Korea
- 김지수Department of Cogno-Mechatronics Engineering, Pusan National University, Busan, 46241, Korea
- 김창석교신저자Department of Cogno-Mechatronics Engineering, Pusan National University, Busan, 46241, Korea
초록
We designed and developed an fNIRS system that suppresses this noise by integrating a short-channel-based signal correction method and frequency division multiplexing (FDM). The results indicated that after applying short-channel regression, the effect size (Cohen’s d) increased by approximately 0.9 compared to the measurements taken before applying short-channel regression. Additionally, a significant difference in hemodynamic responses was observed between the digit span task and the control task. These findings suggest that the proposed fNIRS device is unaffected by ambient light and that short-channel regression effectively reduces noise interference. Consequently, this system can deliver accurate and stable measurements of brain activity, even in conditions where managing ambient light and surface noise is complicated.
키워드
Biomedical sensorBrain imagingfNIRSOptical imagingWearable device