Effectiveness of Radiation Shields to Minimize Operator Dose in the Bronchoscopy Suite: A Phantom Study and Clinical Application

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q1Scopus Q1
저널
JOURNAL OF CLINICAL MEDICINE14(6) · #2114 · 2025
피인용
Web of Science 4Scopus 4
FWCI (Scopus)
1.21
JCR IF
JCR 2024 2.9
ISSN
2077-0383

저자

  1. Jeon, Hosang제1저자Univ Texas Southwestern Med Ctr, Dept Radiat Oncol, Dallas, TX 75390 USA
  2. 김동운Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan 50612, South Korea
  3. 주지현Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan 50612, South Korea
  4. 기용간Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan 50612, South Korea
  5. 강석웅Pusan Natl Univ, Pusan Natl Univ Sch Med, Dept Orthoped, Yangsan Hosp, Yangsan 50612, South Korea
  6. 신원철Pusan Natl Univ, Pusan Natl Univ Sch Med, Dept Orthoped, Yangsan Hosp, Yangsan 50612, South Korea
  7. 윤성훈Pusan Natl Univ, Div Pulm & Crit Care Med, Yangsan Hosp, Yangsan 50612, South Korea
  8. 김윤성Pusan Natl Univ, Div Pulm & Crit Care Med, Yangsan Hosp, Yangsan 50612, South Korea
  9. Yong, Seung HyunYonsei Univ, Severance Hosp, Dept Internal Med, Div Pulm & Crit Care Med,Coll Med, Seoul 03722, South Korea
  10. Chung, Hyun SungNatl Canc Ctr, Dept Internal Med, Div Pulmonol, Goyang 10408, South Korea
  11. Lee, TaehoonUniv Ulsan, Ulsan Univ Hosp, Dept Internal Med, Coll Med, Ulsan 44033, South Korea
  12. 설희윤교신저자Pusan Natl Univ, Div Pulm & Crit Care Med, Yangsan Hosp, Yangsan 50612, South Korea

초록

Background/Objectives: Fluoroscopy has been widely adopted in interventional pulmonology, as it facilitates real-time visualization of the bronchoscope, endobronchial ultrasound, and biopsy tools during procedures. The purpose of this study was to evaluate the effectiveness of radiation shields in minimizing scattered X-ray dose to the bronchoscopist in a phantom study and to determine the dose of scattered X-ray dose to medical staff with radiation shields in clinical application. Methods: An anthropomorphic torso phantom was positioned on the fluoroscopic table between the C-arm X-ray tube and the image detector to mimic bronchoscopic operations. Upper and lower body lead shields were used to examine the effectiveness of radiation shielding. Scatter radiation rates were assessed at a first operator location using real-time dosimeters with and without protective devices. In clinical application, the scattered X-ray dose of the first operator and main assistant was measured using wearable radiation dosimeters during 20 procedures. Results: In the phantom study, scattered radiation without shielding was 266.34 +/- 8.86 mu Sv/h (glabella), 483.90 +/- 8.01 mu Sv/h (upper thorax), 143.97 +/- 8.20 mu Sv/h (hypogastrium), and 7.22 +/- 0.28 mu Sv/h (ankle). The combination of upper and lower body lead shields reduced the scattered X-ray dose by 98.7%, 98.3%, 66.2%, and 79.9% at these levels, respectively. In clinical application, mean scattered X-ray dose rates were 0.14 +/- 0.05 mu Sv/procedure (eye), 0.46 +/- 0.51 mu Sv/procedure (chest), 0.67 +/- 0.50 mu Sv/procedure (hypogastrium), and 1.57 +/- 2.84 mu Sv/procedure (assistant's wrist). Conclusions: The combination of radiation shields significantly reduced the scattered X-ray dose at the operator site in the phantom study. The scattered X-ray dose to medical staff during bronchoscopy can be kept at a low level with the aid of a shielding system.

키워드

scattered radiationradiation shieldsbronchoscopistassistantphantom studyclinical applicationPERIPHERAL PULMONARY-LESIONSSTAFFPROTECTIONEXPOSUREIMPACTTIME