본문 바로가기

Effect of electrochemical topology on detection sensitivity in MEA assay for drug-induced cardiotoxicity screening

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q1Scopus Q1
피인용
Web of Science 2Scopus 1
FWCI (Scopus)
0.21
JCR IF
JCR 2024 10.5
ISSN
0956-5663

저자

  1. Kim, Byunggik제1저자Johns Hopkins Univ, Dept Mech Engn, Baltimore, MD 21218 USA
  2. Choi, Jong SeobKongju Natl Univ, Div Adv Mat Engn, Budaedong 275, Cheonan 31080, Chungnam, South Korea
  3. Zhu, YiguangJohns Hopkins Univ, Dept Environm Hlth & Engn, Baltimore, MD 21205 USA
  4. Kim, JuhyunKorea Brain Res Inst KBRI, Emot Cognit & Behav Res Grp, Daegu 41062, South Korea
  5. 김예슬Pusan Natl Univ, Sch Med, Dept Physiol, Yangsan 50612, South Korea
  6. Parra, AndresJohns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
  7. Locke, Paul A.Johns Hopkins Univ, Dept Environm Hlth & Engn, Baltimore, MD 21205 USA
  8. 김재호Pusan Natl Univ, Sch Med, Dept Physiol, Yangsan 50612, South Korea
  9. Herron, ToddUniv Michigan, Dept Internal Med, Ann Arbor, MI 48109 USA
  10. Kim, Deok-Ho교신저자Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA

초록

Cardiotoxicity remains a major challenge in drug development, accounting for 45% of medication withdrawals due to cardiac ischemia and arrhythmogenicity. To overcome the limitations of traditional multielectrode array (MEA)-based cardiotoxicity assays, we developed a Nafion-coated NanoMEA platform with decoupled reference electrodes, offering enhanced sensitivity for electrophysiological measurements. The 'Decoupled' configuration significantly reduced polarization resistance (Rp) from 12.77 MΩ to 3.41 MΩ, improving charge transfer efficiency as demonstrated by electrochemical impedance spectroscopy and cyclic voltammetry. Additionally, the limit of detection significantly decreased from 0.175 MΩ (Coupled) to 0.040 MΩ (Decoupled), underscoring the system's enhanced sensitivity. Using human-induced pluripotent stem cell-derived cardiomyocytes (hiPSC-CMs), we evaluated the effects of three proarrhythmic drugs: Ranolazine, Domperidone, and Sotalol. Under the decoupled condition, the platform exhibited reductions in IC50 values for Domperidone (0.71 μM–0.29 μM), Sotalol (7.61 μM–0.27 μM), and Ranolazine (53.08 μM–5.89 μM), demonstrating significantly improved drug detection sensitivity. Longitudinal analysis revealed significant alterations in key electrophysiological parameters, including beating period (BP), field potential duration (FPD), spike slope, and amplitude, which were consistent with the known pharmacological actions of these drugs. Further validation through action potential (AP) waveform analysis showed enhanced repolarization dynamics, confirming the platform's predictive capabilities. Our findings highlight the critical role of electrochemical topology in optimizing MEA performance. The NanoMEA system, featuring decoupled Nafion-coated electrodes, represents a robust and sensitive platform for cardiotoxicity screening, setting a new standard for preclinical drug safety assessment and advancing bioelectronic device design for cardiac research.

키워드

Cardiotoxicity screeningElectrochemical topologyHuman iPSC-derived cardiomyocytesImpedance spectroscopyMicroelectrode array (MEA)NanoMEA platformCELL-DERIVED CARDIOMYOCYTESHIGH-THROUGHPUTHERG