영상 딥러닝 기반 소변주머니 내부 소변량 자동 감시 기술 개발
Development of An Automated Liquid Level Monitoring Technique for Urine Bag Based on Deep Learning
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCI
- 저널 순위
- KCI Q4
- 저널
- 의공학회지45(6) · pp.330 · 2024
- 피인용
- KCI 0
- ISSN
- 1229-0807
저자
- 김민재제1저자부산대학교 의학연구원
- 수브라타 바타차르지Medical Research Institute, Pusan National University, Yangsan, Korea
- 김건호양산부산대학교병원 의공학과
- 남경원교신저자양산부산대학교병원 의공학과,부산대학교 의과대학 의공학교실
초록
In-hospital patients who need long-term catheterized urinary support requests repetitive flushing of urine in the urine bag, which can increase the workload of nursing staffs. In this study, a deep learning-based liquid level monitoring technique (based on EfficientDet Lite models) that can estimate the status of in-bag liquid level in real- time from the periodically photographed bag images captured using a camera installed in front of the urine bag. In experiments, the error rates between the manually-calculated in-bag liquid area and the model-extracted in-bag liquid area were 1.05 ± 1.58%, and those between the manually-calculated scale area and the model-extracted scale area were 2.57 ± 2.54%, respectively. In addition, the error rates between the actual amount of in-bag liquid and the model-calculated in-bag liquid amount were 2.28 ± 4.97%, respectively. The implemented technique could success- fully estimate the time-varying amount of liquid in the urine bag. It can be applied to develop a smart flush-request alarm system for patients with long-term catheterized urinary support that can reduce the workload of nursing staffs while increasing the safety of patients.
키워드
Catheterized urinary supportUrine bagDeep learningSmart alarm
식별자
- KCI
- ART003151811



로그인
연구물 목록으로