본문 바로가기

Development and evaluation of modified lead gloves to reduce hand radiation dose during interventional radiological procedures

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q1Scopus Q1KCI Q2
저널
NUCLEAR ENGINEERING AND TECHNOLOGY56(7) · pp.2781 · 2024
피인용
Web of Science 2Scopus 2KCI 0
FWCI (Scopus)
0.26
JCR IF
JCR 2023 2.6
ISSN
1738-5733

저자

  1. 박현준제1저자Busan Natl Univ, Yangsan Hosp, Dept Radiol, Yangsan 50612, South Korea
  2. Jo, ByungduDongseo Univ, Dept Radiol Sci, Busan 47011, South Korea
  3. Lee, Seung-Jae교신저자Dongseo Univ, Dept Radiol Sci, Busan 47011, South Korea

초록

We developed lead gloves that minimize radiation dose to the operator's hands during interventional radiological procedures and that do not impede the operator's surgical capabilities. Existing lead gloves can protect the operator's hands by shielding radiation, but use of such gloves may impair preception sensitivity, resulting in a reduction in the operator's surgical ability. Accordingly, in this study, we developed modified lead gloves that can reduce radiation dose while maintaining operator sensitivity during procedures by modifying the operator's main surgical finger area in existing lead gloves. To evaluate the performance of developed modified lead gloves, radiation was applied in surgical conditions without gloves and with surgical gloves, lead gloves, and modified lead gloves. The radiation dose was evaluated for each condition. When the modified lead gloves were worn, the degree of shielding was similar to when conventional lead gloves were worn. Based on these results, if the operator wears modified lead gloves during interventional radiological procedures, they will protect the hands from radiation while maintaining physical sensitivity in the hands.

키워드

InterventionLead gloveRadiation doseEXPOSUREPATIENTSTAFF

식별자

Scopus EID
2-s2.0-85186325216
WoS UID
001258073200001
KCI
ART003088641