Repeatability and Reliability Evaluation of a Wireless Head-band Sensor

문서 유형
연구출판물 · 학술지 논문
등재
KCI
저널
운동학 학술지 (운동학 학술지)20(4) · pp.70 · 2018
피인용
KCI 0
ISSN
2586-5595

저자

  1. Stephen Tiernan제1저자Dublin Institute of Technology Tallaght, Dublin, Republic of Ireland
  2. 데이비드 오설리번교신저자Pusan National University, Busan, Republic of Korea
  3. Gary ByrneDublin Institute of Technology Tallaght, Dublin, Republic of Ireland

초록

[OBJECTIVES] The objective of this study to examine the reliability and repeatability of the headband sensor in comparison with the gold standard 3 linear and 3 angular rate sensors placed at the center of gravity of the Hybrid III dummy head during drop tests. [METHODS] A SIM-G headband sensor was attached to a Hybrid III dummy head and neck which was equipped with a triaxial accelerometer and 3 angular rate sensors. Linear acceleration and angular rate was sampled at 10,000 Hz while the SIM-G headband was sampled at 1000 Hz and 800 Hz respectively. A drop test was developed to test between 20 to 140 g and consisted of a total of 400 impacts in four locations, left occipital, right occipital, frontal and rear impacts with 100 impacts per location. Multiple tests were performed at the same height to verify the reliability of the devices. SIM-G data and the reference sensors were compared for validity. [RESULTS] The reliability measured by Cronbach's alpha showed very high repeatability for the SIM-G (α = 0.97-0.99). However, the validity measurement, Pearson's Correlation coefficient (r), showed a weak to very strong relationship (r=0.2-0.9). [CONCLUSIONS] While using the SIM-G sensor to help monitor head impact, the weak to strong validity of the SIM-G sensor must be carefully considered by the clinicians or researchers as a serious limitation.

식별자

KCI
ART002402728