Comparison of automated quantification of amyloid deposition between PMOD and Heuron
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q2Scopus Q1
- 저널
- SCIENTIFIC REPORTS13(1) · 2023
- 피인용
- Web of Science 10Scopus 11
- FWCI (Scopus)
- 0.56
- JCR IF
- JCR 2022 4.6
- ISSN
- 2045-2322
저자
- Roh, Hyun Woong제1저자Ajou Univ, Sch Med, Dept Psychiat, Suwon, South Korea
- Son, Sang JoonAjou Univ, Sch Med, Dept Psychiat, Suwon, South Korea
- Hong, Chang HyungAjou Univ, Sch Med, Dept Psychiat, Suwon, South Korea
- Moon, So YoungAjou Univ, Sch Med, Dept Neurol, Suwon, South Korea
- Lee, Sun MinAjou Univ, Sch Med, Dept Neurol, Suwon, South Korea
- Seo, Sang WonSungkyunkwan Univ, Samsung Med Ctr, Sch Med, Dept Neurol, Seoul, South Korea
- Choi, Seong HyeInha Univ, Sch Med, Dept Neurol, Incheon, South Korea
- 김은주Pusan Natl Univ, Pusan Natl Univ Hosp, Med Res Inst, Dept Neurol,Sch Med, Busan, South Korea
- Cho, Soo HyunChonnam Natl Univ, Chonnam Natl Univ Hosp, Dept Neurol, Med Sch, Gwangju, South Korea
- Kim, Byeong ChaeChonnam Natl Univ, Chonnam Natl Univ Hosp, Dept Neurol, Med Sch, Gwangju, South Korea
- Park, SeongbeomHeuron Co Ltd, Incheon, South Korea
- Song, SoohwaHeuron Co Ltd, Incheon, South Korea
- An, Young-Sil교신저자Ajou Univ, Sch Med, Dept Nucl Med & Mol Imaging, 206 World Cup Ro, Suwon 16499, Gyeonggi Do, South Korea
초록
Several programs are widely used for clinical and research purposes to automatically quantify the degree of amyloid deposition in the brain using positron emission tomography (PET) images. Given that very few studies have investigated the use of Heuron, a PET image quantification software approved for clinical use, this study aimed to compare amyloid deposition values quantified from 18F-flutemetamol PET images using PMOD and Heuron. Amyloid PET data obtained from 408 patients were analysed using each quantitative program; moreover, the standardized uptake value ratios (SUVRs) of target areas were obtained by dividing the standardized uptake value (SUV) of the target region by the SUV of cerebellar grey matter as a reference. Compared with PMOD, Heuron yielded significantly higher SUVRs for all target areas (paired sample t-test, p < 0.001), except for the PC/PCC (p = 0.986). However, the Bland–Altman plot analysis indicated that the two quantitative methods may be used interchangeably. Moreover, receiver operating characteristic curve analysis revealed no significant between-method difference in the performance of the SUVRs in evaluating the visual positivity of amyloid deposits (p = 0.948). In conclusion, Heuron and PMOD have comparable performance in quantifying the degree of amyloid deposits in PET images.
키워드
식별자
- Scopus EID
- 2-s2.0-85162223635
- WoS UID
- 001015400200023



로그인
연구물 목록으로