Intensity-modulated dental CBCT: Noise and absorbed dose study

문서 유형
연구출판물 · 학술대회 발표논문
등재
Scopus
저널 순위
Scopus Q3
피인용
Scopus 0
FWCI (Scopus)
0
ISSN
1605-7422

저자

  1. 김진우제1저자School of Mechanical Engineering, Pusan National University, Geumjeong-gu, Busan, South Korea
  2. 윤종희School of Mechanical Engineering, Pusan National University, Geumjeong-gu, Busan, South Korea
  3. 하승우School of Mechanical Engineering, Pusan National University, Geumjeong-gu, Busan, South Korea
  4. 윤한빈Center for Advanced Medical Engineering, Pusan National University, Geumjeong-gu, Busan, South Korea
  5. 김호경교신저자School of Mechanical Engineering, Pusan National University, Geumjeong-gu, Busan, South Korea

초록

Dental computed tomography (CT) typically uses a cone-beam geometry with a flat-panel detector. Although the flat-panel detector normally covers the maxillary and mandibular jaws, the cone-beam scan can deliver the dose to organs that are located out of the direct beam path but sensitive to radiation damage, such as eye lens. For typical dentoalveolar cone-beam CT (CBCT) scans, this study investigates the absorbed dose distributions in the head and neck using the Monte Carlo technique, and quantifies specific organ doses. Then, we design an intensity-modulated CBCT scan protocol that can provide a higher tomographic image quality at a lesser patient dose. The beam-intensity modulation includes the changes of tube current (mA) and/or voltage (kVp) during circular scanning, and the modulation scenarios are designed considering the cervical spine through which x-ray beam attenuates largely. We assess the noise-to-dose performances for various intensity-modulation scenarios, and compare the results with that obtained for the conventional scan.

키워드

absorbed dosedental CTimage qualityintensity modulation