본문 바로가기

Effect from Autoclave Sterilization and Usage on the Fracture Resistance of Heat-Treated Nickel-Titanium Rotary Files

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q2
저널
MATERIALS16(6) · #2261 · 2023
피인용
Web of Science 8Scopus 11
FWCI (Scopus)
0.79
JCR IF
JCR 2022 3.4
ISSN
1996-1944

저자

  1. El Abed, Rashid제1저자Mohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Endodont Dept, Dubai 505055, U Arab Emirates
  2. Al Raeesi, DanaEmirates Hlth Serv Estab, Endodont Dept, Dubai, U Arab Emirates
  3. Alshehhi, AishaMohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Endodont Dept, Dubai 505055, U Arab Emirates
  4. Alkhatib, ZuhairMohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Endodont Dept, Dubai 505055, U Arab Emirates
  5. Khamis, Amar H.Mohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Biostat Dept, Dubai 505055, U Arab Emirates
  6. Jamal, MohamedMohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Endodont Dept, Dubai 505055, U Arab Emirates
  7. 김현철교신저자Pusan Natl Univ, Dent Res Inst, Sch Dent, Dept Conservat Dent, Yangsan 50612, South Korea

초록

This study aimed to assess the effect of mechanical loading and heating on the cyclic fatigue and torsional fracture resistances of heat-treated nickel-titanium files after usage and autoclaving. Sixty files (One Curve) were tested for cyclic fatigue and torsional fracture resistances using customized devices. The files were divided into three groups according to the test conditions (n = 10); new (group-N), used for simulated canal shaping (group-U), and sterilized after use (group-S). For cyclic fatigue resistances, the files were freely rotated in a curved metal canal under body temperature; the time elapsed to fracture was recorded and the numbers of cycles to fracture (NCF) were calculated. For the torsional resistances, the file tip was fixed and rotated until the file fractured. The maximum torsional load and distortion angle were recorded. The toughness was calculated. Fracture fragments were examined with a scanning electron microscope. Data were analyzed using a one-way analysis of variance and Tukey's post hoc test at the significance level of 95%. Group-U showed significantly higher NCF than group-S (p < 0.05). However, there was no significant differences between groups-N and -S in the NCF (p > 0.05). Group-N showed a significantly bigger distortional angle and higher torsional toughness than groups-U and -S, but the ultimate torsional strength did not have significant difference between the groups. Under the limitation of this study, autoclave sterilization after single-usage did not improve the fracture resistance of heat-treated One Curve nickel-titanium files.

키워드

autoclave sterilizationdynamic fatiguefracture resistanceheat treatmentnickel-titanium filetorsional strengthCYCLIC FATIGUE RESISTANCEMECHANICAL-PROPERTIESTORSIONAL RESISTANCEBENDING PROPERTIESSHAPE-MEMORYINSTRUMENTS

식별자

Scopus EID
2-s2.0-85151504438
WoS UID
000959081300001