2D 영상마커 추적 기반 시린지펌프 투약속도 실시간 감시 기술 개발
Real-time Monitoring of the Actual Infusion Rate of Syringe Pump Using 2D Image Marker Tracking
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCI
- 저널 순위
- KCI Q4
- 저널
- 의공학회지44(1) · pp.92 · 2023
- 피인용
- KCI 2
- ISSN
- 1229-0807
저자
초록
Purpose: To propose a new infusion rate monitoring technique based on the 2D image marker tacking to improve patient safety by preventing syringe pump-related medication accidents due to decreased infusion rate control accuracy. Materials and Methods: The infusion rate of the syringe pump and drug residue in the pump-equipped syringe were monitored in real time by tracking the movement of the 2D image markers attached to the syringe pump. Results: The error rate between the set and the estimated infusion rates was 1.03, 0.66, 1.95, 0.23, and 1.05% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. In addition, the error rate between the actual and the estimated drug residues was 1.04, 0.47, 0.60, 3.66, and 0.00% when the infusion rate setting was 10, 20, 30, 40, and 50 mL/H, respectively. Conclusion: Experimental results demonstrated that the proposed technique can increase the efficiency of the safety management system for seriously ill inpatients by decreasing a possibility of syringe pump- related medication accidents in hospitals.
키워드
syringe pumppatient safety2D image markerreal-time monitoring