A Study on Radiation Dose and Image Quality using High-Pitch Mode when Examining the Lumbar Region with Dual-Energy Computed Tomography

문서 유형
연구출판물 · 학술지 논문
등재
KCI
저널 순위
KCI Q4
저널
International Journal of Contents18(4) · pp.23 · 2022
피인용
KCI 0
ISSN
1738-6764

저자

  1. 김경립제1저자양산부산대학교병원
  2. 곽종혁교신저자양산부산대학교병원
  3. 강효석

초록

The purpose of this study was to investigate whether the high pitch protocol method of dual-energy computed tomography can be successfully applied to a lumbar vertebrae CT scan and reduce the effect of exposure dose; the main indicators were the image quality and reduced scan time. This study was conducted using MDCT 256 row dual source, Somatom Definition Flash (Siemens medical system: Germany), on 69 patients who had lumbar vertebrae CT scans from March 1, 2021, to June 30, 2021. First, to determine the applicability of the high pitch protocol method, the phantom images quality was compared with that of the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol method with the AAPM phantom, respectively. In the phantom image, CT noise, uniformity, and spatial resolution were not significantly different between the existing lumbar vertebrae CT scan method and the examination method using the high pitch protocol. In addition, in images examined from clinical patients, single-sample ttest results of the comparison of the mean value of CT noise, uniformity, and spatial resolution did not show statistically significant differences, 0.081, 0.078, and 0.387, respectively. The results of the qualitative evaluation showed that all were also higher than 3 points, which is fair, and the radiation dose was reduced by 9.76% in the high pitch protocol. When examining the lumbar region with dual-energy computed tomography, the examination method using high-pitch mode reduced the radiation dose and shortened the examination time, with the same image quality as that of the existing lumbar vertebrae CT scan.

키워드

Dual-Energy Computed TomographyLumbar VertebraeHigh Pitch ProtocolRadiation Dose

식별자

KCI
ART002920222