본문 바로가기

Early unexpected failure of a vitamin E-infused highly cross-linked polyethylene liner A case report

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q3Scopus Q2
저널
MEDICINE100(41) · #e27454 · 2021
피인용
Web of Science 5Scopus 5
FWCI (Scopus)
0.35
JCR IF
JCR 2020 1.889
ISSN
0025-7974

저자

  1. 김경백제1저자Pusan Natl Univ, Dept Orthopaed Surg, Yangsan Hosp, Yangsan, South Korea
  2. 이상민Pusan Natl Univ, Res Inst Convergence Biomed Sci & Technol, Sch Med, Yangsan, South Korea
  3. 문남훈Pusan Natl Univ, Res Inst Convergence Biomed Sci & Technol, Sch Med, Yangsan, South Korea
  4. 도민욱Pusan Natl Univ, Dept Orthopaed Surg, Yangsan Hosp, Yangsan, South Korea
  5. 신원철교신저자Pusan Natl Univ, Res Inst Convergence Biomed Sci & Technol, Sch Med, Yangsan, South Korea

초록

Rationale: Total hip arthroplasty (THA) with a polyethylene (PE) liner is 1 of the most effective and successful treatment strategies for end-stage hip disorders. Vitamin E-infused highly cross-linked polyethylene is theoretically known to prevent failure due to oxidative degradation in the body, and is resistant to wear; therefore, successful long-term survival of THA is expected. Patient concerns: In June 2019, approximately 1.5 years after THA, the patient sat down and stood up without any specific issue; however, an unusual bullet sound occurred around the left hip joint. Since then, his discomfort persisted, and he was admitted to the emergency department. Diagnosis: Plain radiography and metal artifact reduction computed tomography performed in the emergency department revealed eccentric elevation of the prosthetic femoral head and suspected PE liner failure. Intervention: Revision surgery was performed for modular component exchange. To reduce the dislocation risk without performing cup exchange, conversion to dual-mobility articulation was performed. Outcomes: During the regular follow-up post-surgery, the patient could perform daily life activities without any discomfort, and dislocation was not observed. At 2 years postoperatively, no significant changes were observed in the radiographic images. Lessons: This case report presents an unexpected failure of THA due to superior rim fracture and excessive wear at the locking mechanism of the vitamin E-infused highly cross-linked polyethylene liner. This is an interesting case, as early PE liner failure occurred without strong labor intensity or trauma. A modular component exchange was performed with a dual-mobility bearing, and no issues were observed approximately 2 years after the reoperation. Therefore, third-generation highly cross-linked polyethylene liners can also cause early failure without a clear cause, and this case report highlights the necessity of considering several strategies for reoperation. Study Design: Case report.

키워드

dual-mobilitylinerpolyethylenerim fracturevitamin E-infusedwearMOLECULAR-WEIGHT POLYETHYLENETOTAL JOINT ARTHROPLASTYIN-VIVO WEARRESISTANCEOXIDATIONRATES

식별자

Scopus EID
2-s2.0-85121580495
WoS UID
000731129600047