A Miniaturized 256-Channel Neural Recording Interface With Area-Efficient Hybrid Integration of Flexible Probes and CMOS Integrated Circuits

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q1
저널
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING69(1) · pp.334 · 2022
피인용
Web of Science 36Scopus 40
FWCI (Scopus)
1.9
JCR IF
JCR 2021 4.756
ISSN
0018-9294

저자

  1. 박성윤제1저자Pusan Natl Univ, Dept Elect Engn, Busan, South Korea
  2. Na, KyounghwanUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA
  3. Voroslakos, MihalyUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA
  4. Song, HyunsooUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA
  5. Slager, NathanUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA
  6. Oh, SungjinUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA
  7. Seymour, JohnUniv Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst WIMS, Ann Arbor, MI 48109 USA
  8. Buzsaki, GyorgyNYU, New York, NY 10003 USA
  9. Yoon, Euisik교신저자Univ Michigan, Dept Elect Engn & Comp Sci, Ctr Wireless Integrated Micro Sensing & Syst, Ann Arbor, MI 48109 USA

초록

We report a miniaturized, minimally invasive high-density neural recording interface that occupies only a 1.53 mm(2) footprint for hybrid integration of a flexible probe and a 256-channel integrated circuit chip. To achieve such a compact form factor, we developed a custom flip-chip bonding technique using anisotropic conductive film and analog circuit-under-pad in a tiny pitch of 75 mu m. To enhance signal-to-noise ratios, we applied a reference-replica topology that can provide the matched input impedance for signal and reference paths in low-noise aimpliers (LNAs). The analog front-end (AFE) consists of LNAs, buffers, programmable gain amplifiers, 10b ADCs, a reference generator, a digital controller, and serial-peripheral interfaces (SPIs). The AFE consumes 51.92 mu W from 1.2 V and 1.8 V supplies in an area of 0.0161 mm(2) per channel, implemented in a 180 nm CMOS process. The AFE shows > 60 dB mid-band CMRR, 6.32 mu V-rms input-referred noise from 0.5 Hz to 10 kHz, and 48 M omega input impedance at 1 kHz. The fabricated AFE chip was directly flip-chip bonded with a 256-channel flexible polyimide neural probe and assembled in a tiny head-stage PCB. Full functionalities of the fabricated 256-channel interface were validated in both in vitro and in vivo experiments, demonstrating the presented hybrid neural recording interface is suitable for various neuroscience studies in the quest of large scale, miniaturized recording systems.

키워드

ProbesBondingIntegrated circuit interconnectionsFlip-chip devicesPolyimidesCMOS integrated circuitsMicroelectrodesAnalog circuit-under-padanisotropic conductive filmCMOS integrated circuitelectrophysiologyneural interfacehybrid integrationminimally invasivereference-replica techniqueimpedance bufferingMICROELECTRODE ARRAYFRONT-ENDELECTRODESCOMPACT