MONTE CARLO DOSE ASSESSMENT IN DENTAL CONE-BEAM COMPUTED TOMOGRAPHY

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q4Scopus Q3
저널
RADIATION PROTECTION DOSIMETRY193(3-4) · pp.190 · 2021
피인용
Web of Science 3Scopus 4
FWCI (Scopus)
0.22
JCR IF
JCR 2020 0.972
ISSN
0144-8420

저자

  1. 김진우제1저자Pusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea
  2. 전호상Pusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan Si 50612, Gyeongsangnam D, South Korea
  3. 김호경교신저자Pusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea

초록

Most dental cone-beam computed tomography (CBCT) uses an x-ray beam field covering the maxillomandibular region and the width-truncated detector geometry. The spatial dose distribution in dental CBCT is analyzed in terms of local primary and remote secondary doses by using a list-mode analysis of x-ray interactions obtained from the Monte Carlo simulations. The patient-dose benefit due to the width-truncated detector geometry is also investigated for a wide range of detector offsets. The developed dose estimation agrees with the measurement in a relative error of 7.7%. The secondary dose outside of the irradiation field becomes larger with increasing tube voltage. The dose benefit with the width-truncated geometry linearly increases as the detector-offset width is decreased. Leaving the CT image quality out of the account, the MC results reveal that the operation of dental CBCT with a lower tube voltage and a smaller detector-offset width is beneficial to the patient dose.

키워드

TUBE CURRENT MODULATIONIMAGE QUALITYCHEST CTCBCT DEVICESREDUCTIONDOSIMETRYRECONSTRUCTIONOPTIMIZATIONSCANNERSPHANTOM

식별자

Scopus EID
2-s2.0-85106554813
WoS UID
000707114400008