본문 바로가기

단일 광섬유 센서 프로브의 국소 기포 변수 측정 성능 평가

Evaluation of Local Bubble Parameter Measurement Performance of Single Sensor Optical Fiber Probe

문서 유형
연구출판물 · 학술지 논문
등재
KCI
저널 순위
KCI Q2
저널
한국유체기계학회 논문집24(1) · pp.27 · 2021
피인용
KCI 1
ISSN
2287-9706

저자

  1. 문정민제1저자부산대학교
  2. 김태호부산대학교
  3. 박진영부산대학교
  4. 윤병조교신저자부산대학교

초록

Bubble parameters such as the void fraction, bubble velocity, and bubble diameter are important for an accurate prediction of bubble behaviors in two-phase flow. In recent decades, a number of researchers have developed a single optical fiber sensor probe to measure the local bubble parameters and verified its performance in a velocity range up to 1.5m/s. However, further validation of a single fiber sensor probe over a wider range of flow conditions is necessary to achieve its general applicability. In this study, the performance of a single sensor probe was evaluated through air-water experiments. Also, its measurement performance was compared with that of a 2-sensor optical fiber sensor probe. From the test, it was found that the velocity measurement error of a single fiber optical sensor probe was about 4.8% in the velocity range from 0.2m/s to 4m/s. In addition to this, it was confirmed that the single optical fiber sensor is able to measure the bubble diameter within the 26.2% error range.

키워드

Optical fiber sensor(광섬유 센서)Local bubble parameter(국소 기포 변수)Single sensor probe(단일 센서 프로브)Double sensor probe(2중 센서 프로브)Optical fiber sensorLocal bubble parameterSingle sensor probeDouble sensor probe

식별자

KCI
ART002682765