단일 광섬유 센서 프로브의 국소 기포 변수 측정 성능 평가
Evaluation of Local Bubble Parameter Measurement Performance of Single Sensor Optical Fiber Probe
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCI
- 저널 순위
- KCI Q2
- 저널
- 한국유체기계학회 논문집24(1) · pp.27 · 2021
- 피인용
- KCI 1
- ISSN
- 2287-9706
저자
초록
Bubble parameters such as the void fraction, bubble velocity, and bubble diameter are important for an accurate prediction of bubble behaviors in two-phase flow. In recent decades, a number of researchers have developed a single optical fiber sensor probe to measure the local bubble parameters and verified its performance in a velocity range up to 1.5m/s. However, further validation of a single fiber sensor probe over a wider range of flow conditions is necessary to achieve its general applicability. In this study, the performance of a single sensor probe was evaluated through air-water experiments. Also, its measurement performance was compared with that of a 2-sensor optical fiber sensor probe. From the test, it was found that the velocity measurement error of a single fiber optical sensor probe was about 4.8% in the velocity range from 0.2m/s to 4m/s. In addition to this, it was confirmed that the single optical fiber sensor is able to measure the bubble diameter within the 26.2% error range.
키워드
Optical fiber sensor(광섬유 센서)Local bubble parameter(국소 기포 변수)Single sensor probe(단일 센서 프로브)Double sensor probe(2중 센서 프로브)Optical fiber sensorLocal bubble parameterSingle sensor probeDouble sensor probe