Ipsilateral ultrasound-monitoring technique for reducing malpositions of peripherally inserted central catheters in the intensive care unit
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q3Scopus Q2
- 저널
- ANNALS OF PALLIATIVE MEDICINE10(2) · pp.1530 · 2021
- 피인용
- Web of Science 2Scopus 3
- FWCI (Scopus)
- 0.12
- JCR IF
- JCR 2020 2.595
- ISSN
- 2224-5820
저자
- 송승환제1저자Department of Thoracic and Cardiovascular Surgery, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 허업교신저자Department of Thoracic and Cardiovascular Surgery, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 이재일Department of Neurosurgery, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 이충원Department of Thoracic and Cardiovascular Surgery, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 엄중섭Division of Respiratory, Department of Internal Medicine, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 김효정Division of Hematology-Oncology, Department of Internal Medicine, Medical Research Institute, Pusan National University Hospital, School of Medicine, Busan, Republic of Korea
- 왕일재Department of Emergency Medicine, Biomedical Research Institute, Pusan National University Hospital, Pusan National University School of Medicine, Busan, Republic of Korea
- 김재준Division of Hematology-Oncology, Department of Internal Medicine, Medical Research Institute, Pusan National University Yangsan Hospital, School of Medicine, Yangsan, Republic of Korea
초록
Background: We performed ultrasonographic examinations of the ipsilateral internal jugular vein after placement of peripherally inserted central catheters (PICCs) and found that more than half of malpositions of PICC went to the head through the ipsilateral internal jugular vein. We aimed to reduce malpositions of PICCs in settings where fluoroscopy is not available. Methods: This retrospective study had a case-control, single-center study design. From January to June 2018, PICCs were inserted with the patient's arm abducted without the use of ultrasonography to identify the presence of the PICC in the ipsilateral internal jugular vein (procedure A). From July to December 2018, PICCs were inserted in a similar fashion; however, ultrasonography was employed to detect the catheter tip in the ipsilateral internal jugular vein (procedure B). Results: There were 794 placements of PICCs in 650 patients. Procedures A and B were performed in 418 and 376 patients, respectively. Malpositioning of PICC tips occurred significantly less often in procedure B than in procedure A (2.4% vs. 5.7%, P=0.02). Conclusions: In a setting without fluoroscopy, insertions of PICCs can be performed with the arm abducted at 90 degrees. Ultrasound monitoring for the presence of the catheter in the ipsilateral internal jugular vein can help with correct positioning, thereby reducing the occurrence of catheter tip malpositions.
키워드
Malpositionperipherally inserted central catheter (PICC)ultrasoundCENTRAL VENOUS CATHETERBLOOD-STREAM INFECTIONSLARGE VEIN-THROMBOSISTIPPLACEMENTRISK