본문 바로가기

Implant placement in the removable mandibular advancement device for completely edentulous patients: a finite element study

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q1Scopus Q1KCI Q1
피인용
Web of Science 15Scopus 15KCI 4
FWCI (Scopus)
1.13
JCR IF
JCR 2020 5.86
ISSN
2288-4300

저자

  1. Park, Jieun제1저자School of Mechanical Engineering, Kyungpook National University, Daegu 41566, South Korea
  2. Ahn, Su-JinDepartment of Biomaterials and Prosthodontics, Kyung Hee University Hospital at Gangdong, School of Dentistry, Kyung Hee University, Seoul 05278, South Korea
  3. 이현종교신저자Department of Prosthodontics, Dental Research Institute, Dental and Life Science Institute, School of Dentistry, Pusan National University, Yangsan 50612, South Korea
  4. Noh, Gunwoo교신저자School of Mechanical Engineering, Kyungpook National University, Daegu 41566, South Korea

초록

Few studies have examined the biomechanical effects of dental implant placement in the implant-supported mandibular advancement device (MAD) for the treatment of obstructive sleep apnea in completely edentulous patients. This biomechanical study aimed to evaluate the stability of the implant-supported MAD for different numbers and positions of dental implants. A 3D edentulous maxillofacial model with the one-piece MAD was considered for various numbers and positions of implants. On each mandible and maxilla, one, two, four, or six implants were placed symmetrically in various positions. Total 42 implant placements were considered. A static finite element analysis was performed under the mandibular posterior restorative force for 40% protrusion. The maximum and minimum principal stresses on the cortical and cancellous bones and the von Mises stresses on the implant were measured. When four implants were placed on each maxilla and mandible, the maximum average stress on the cancellous bone was reduced by up to approximately three times compared with the cases of placing two implants on each maxilla and mandible. The placement of at least four implants on each maxilla and mandible improved MAD stability. It is noteworthy that the mandibular implants should be arranged adjacently to avoid stress concentration in the second molar.

키워드

mandibular advancement devicemedical deviceimplant-supported dental prosthesisdental implantobstructive sleep apneastress analysisOBSTRUCTIVE SLEEP-APNEARETAINED OVERDENTURESSTRESS-DISTRIBUTIONORAL APPLIANCESPROSTHESISDESIGNBONE