Performance Improvement of Lead (II) Iodide Dosimeter by Addition of Gadolinium Oxysulfide Powder and Applicability to Radiotherapy Quality Assurance Evaluation

문서 유형
연구출판물 · 학술지 논문
등재
SCIE
저널 순위
JCR Q4
저널
SCIENCE OF ADVANCED MATERIALS12(11) · pp.1601 · 2020
피인용
Web of Science 3
JCR IF
JCR 2019 1.117
ISSN
1947-2935

저자

  1. Shin, Yo-Han제1저자인제대학교, Coll Med, Dept Radiat Oncol, KO
  2. Yang, Seung-Woo인제대학교, Coll Med, Dept Radiat Oncol, KO
  3. Han, Moo-Jae인제대학교, Coll Med, Dept Radiat Oncol, KO
  4. 정재훈부산대학교, 부속연구소, 양산병원 의생명융합연구소, KO
  5. Cho, Sam-Joo이화여자대학교, Mokdong Hosp, Dept Radiat Oncol, KO
  6. Kang, Young-Nam가톨릭대학교, Coll Med, Seoul St Marys Hosp, KO
  7. Kim, Ki-Hwan충남대학교, Chungnam Natl Univ Hosp, Dept Radiat Oncol, KO
  8. Cho, Heung-Lae인제대학교, Busan Paik Hosp, Dept Radiat Oncol, KO
  9. Park, Sung-Kwang교신저자인제대학교, Busan Paik Hosp, Dept Radiat Oncol, KO

초록

In radiotherapy, quality assurance (QA) procedures to verify the safety of treatment are considered to be critically important. However, owing to general problems associated with dosimeters, active research has been conducted to replace them. Herein, we intended to maximize the sensitivity of PbI2 on radiation by the visible light emitted from the phosphor. For this purpose, the PbI2 based pure and hybrid sensors containing Gd2O2S:Tb (0-3 wt%), a phosphor with excellent fluorescence efficiency, were fabricated. The sensors were irradiated with photons of 6 MV and 15 MV, and the reaction characteristics were compared and evaluated. Reproducibility evaluation results showed that the criterion of RSD < 1.5% was not met in any of the cases except for the pure PbI2 sensor. In the linearity evaluation, all sensors showed excellent reliability (R-2 = 0.9999). For sensitivity evaluation, the 1 wt% sensor showed values that were 1.88 times and 1.75 times better, respectively for each energy than those of the pure PbI2 sensor, indicating superior performance. The 2 wt% and 3 wt% sensors also showed sensitivity values exceeding 1.13 times that of pure PbI2 sensor, although these values were lower than those of the 1 wt% sensor. None of the sensors showed any significant difference from the pure PbI2 sensor for dose rate dependence and percentage depth dose evaluation.

키워드

Radiation TherapyQuality AssurancePbI2Gd2O2S:TbDosimeter