Performance evaluation of the ZL 6000i cone-plate rotational viscometer

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q4Scopus Q2
저널
CLINICAL HEMORHEOLOGY AND MICROCIRCULATION75(2) · pp.143 · 2020
피인용
Web of Science 1Scopus 1
FWCI (Scopus)
0
JCR IF
JCR 2019 1.741
ISSN
1386-0291

저자

  1. Lee, Ja Young제1저자교신저자Busan Paik Hospital, Inje University, College of Medicine, Busan, South Korea
  2. 오승환Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan, South Korea
  3. Kim, Hye RanBusan Paik Hospital, Inje University, College of Medicine, Busan, South Korea

초록

BACKGROUND: Elevated whole blood viscosity (WBV) is associated with increasing blood flow resistance in the micro-circulation and is used to predict the occurrence and outcome of cardiovascular diseases. OBJECTIVE: We evaluated the analytical performance of the ZL 6000i (Zonci Technology, Beijing, China), new fully automated cone-plate rotational type viscometer. METHODS: We collected blood samples from 287 adults to establish a reference range. We evaluated total precision for 20 days using quality control (QC) samples and within-run precision was evaluated by performing 50 consecutive measurements of blood samples. Aliquots of blood samples were stored at room temperature and 5 degrees C to assess the effect of storage time and temperature on viscosity measurement. RESULTS: The reference ranges and median values of WBV were significantly higher in males than in females due to difference in hematocrit level. The coefficients of variation (CVs) for precision using QC and human blood samples were < 7.5% at all shear rates. WBV of the samples stored at room temperature for up to 6 h, and at 5 degrees C for samples stored up to 2 days was not different from the control. CONCLUSIONS: In this study, the ZL 6000i viscometer yields reproducible WBV data and is clinically useful for monitoring viscosity.

키워드

Blood viscosityviscometervalidationreference valuesstabilityWHOLE-BLOOD VISCOSITYRISKHEMODILUTIONAGGREGATIONRHEOLOGYIMPROVESFITNESS

식별자

Scopus EID
2-s2.0-85089358705
WoS UID
000555842300003