Compact, Portable, High-Density Functional Near-Infrared Spectroscopy System for Brain Imaging

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q1Scopus Q1
저널
IEEE ACCESS8 · pp.128224 · 2020
피인용
Web of Science 28Scopus 33
FWCI (Scopus)
0
JCR IF
JCR 2019 3.745
ISSN
2169-3536

저자

  1. 무하마드 아디프 야쿱제1저자School of Mechanical Engineering, Pusan National University, Busan, South Korea
  2. 우성우School of Mechanical Engineering, Pusan National University, Busan, South Korea
  3. 홍금식교신저자School of Mechanical Engineering, Pusan National University, Busan, South Korea

초록

Advancements in functional near-infrared spectroscopy (fNIRS) neuroimaging have extended its usage from clinical settings to real-life applications. We developed a wearable and wireless fNIRS-based neuroimaging system that can be applied for brain disease diagnostics and during daily life activities. We utilized 128 compact dual-wavelength LEDs of 735 nm and 850 nm and a silicon photodiode (SiPD). The LEDs and SiPD were closely assembled in a 3D-printed flexible configuration holder fitted in a soft neoprene cap. The developed configuration has 128 channels in a 7 x 7 cm area with source-detector separations ranging from 5 mm to 38 mm, which intends to provide a high-density device. The system acquires data from multiple brain depths to generate a 3D activation image. Short-separation channels provide the physiological noise in the superficial layer, which can be used to enhance fNIRS signals. The experimental data are stored in a USB flash disk attached to the device. We developed software that is connected to the device using Wi-Fi to display the oxyhemoglobin (HbO) and deoxyhemoglobin (HbR) levels in real-time. The system architecture allows the selection of channels and the use of multiple devices simultaneously. The system was tested using a silicone phantom as well as human subjects. The LED intensity is dynamically calibrated within the safety range once the device is set up, so that every channel can acquire stable data. The system provided high-quality signals with a signal-to-noise ratio of more than 64 dB, high optical sensitivity, and dynamic optical range of 140 dB in the phantom test. A sampling rate of 30 Hz was achieved with all the channels used. The system exhibited valid HbO and HbR responses during the arterial occlusion test and a motor task. These results show that the developed system can provide precise, robust, and multi-depth measurements resulting in high-density images. With these novel characteristics, the system favors clinical usage and outdoor activities as well as brain-computer interface applications.

키워드

Multi-depth fNIRSbrain-computer interface3D fNIRSwearablewirelessreal-time displayWAVELENGTH COMBINATIONSFNIRSSENSITIVITYIMPROVEMENTHEMOGLOBINTISSUE

식별자

Scopus EID
2-s2.0-85089244287
WoS UID
000551816900001