Dose assessment in dental cone-beam computed tomography: Comparison of optically stimulated luminescence dosimetry with Monte Carlo method

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q1
저널
PLOS ONE15(3) · 2020
피인용
Web of Science 16Scopus 15
FWCI (Scopus)
0
JCR IF
JCR 2019 2.74
ISSN
1932-6203

저자

  1. Lee, Chena제1저자Department of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea
  2. Yoon, JeongminDepartment of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea
  3. Han, Sang-Sun교신저자Department of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea
  4. Na, Ji YeonDepartment of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea
  5. Lee, Jeong-HeeDepartment of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea
  6. Kim, Young HyunDepartment of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea
  7. 황재준Department of Oral and Maxillofacial Radiology, School of Dentistry, Pusan National University, Yangsan, Korea

초록

The variety of cone-beam computed tomography (CBCT) machines and their applications has rapidly increased in recent years, making the dose evaluation of individual devices an important issue. Patient doses from CBCT were assessed with two different methods: optically stimulated luminescence dosimeter (OSLD) measurements and Monte Carlo (MC) simulations, in four different examination modes. Based on an analysis of the measurement process and the obtained values, a recommendation is made regarding which method is more practical and efficient for acquiring the effective dose of CBCT. Twenty-two OSLDs were calibrated and equipped in human phantoms of head and neck organs. They were exposed to radiation from two CBCT units-CS9300 (Carestream Dental LLC, Atlanta, Georgia) and RAYSCAN alpha+ (Ray Co. Ltd, Hwaseong-si, Korea)-using two different examination modes. The dose recorded using the OSLDs was used to calculate the organ dose and the effective dose for each unit in each examination mode. These values were also calculated using MC software, PCXMC (STUK, Helsinki, Finland). The organ doses and effective doses obtained using both methods were compared for each examination mode of the individual units. The OSLD-measured effective dose value was higher than that obtained using the MC method for each examination mode, except the dual jaw mode of CS9300. The percent difference of the effective dose between the two methods ranged from 4.0% to 14.3%. The dose difference between the methods decreased as the field of view became smaller. The organ dose values varied according to the method, although the overall trend was similar for both methods. The organs showing high doses were mostly consistent for both methods. In this study, the effective dose obtained by OSLD measurements and MC simulations were compared, and both methods were described in detail. As a relatively efficient and easy-to-perform method, we cautiously suggest using MC simulations for dose evaluations in the future.