본문 바로가기

Biomechanical Assessment of a Novel Lengthening Plate for Distraction Osteogenesis: A Finite Element Study

문서 유형
연구출판물 · 학술지 논문
등재
KCIScopus
저널 순위
Scopus Q3
저널
Biomedical Engineering Letters (BMEL)6(4) · pp.216 · 2016
피인용
KCI 0
ISSN
2093-9868

저자

  1. 우수헌제1저자Medyssey Co., Ltd
  2. 서승우고려대학교
  3. 정태곤OSONG Medical Innovation Foundation
  4. 이성재인제대학교
  5. 한동욱교신저자부산대학교
  6. 양재혁교신저자고려대학교

초록

Purpose The purpose of this study was to describe a novelbone lengthening plate (BLP) and to evaluate the biomechanicalperformance of the plate before clinical use. Methods A finite element model was obtained from valuesof the compression test. It was used to evaluate the performanceof the BLP, such as biomechanical stability and structuralstiffness, according to materials of the shaft and column usedfor lengthening (type I: all Titanium-alloys, type II and III:Titanium-alloys and Cobalt-Chrome-alloys, type IV: allCobalt-Chrome-alloys). Furthermore, Peak von Mises Stress,risk of screw breakage, and micro-motion between the boneand screw were evaluated. Results Structural stiffness and strain of the finite elementmodel of the femur were similar to cadaveric femurs. Structuralstiffness increased from type I (Titanium alloys) to type IV(Cobalt-Chrome-alloys) by at least 128%. First locking screw,which was inserted at the osteotomy site of the proximalfemur (proximal to distal direction), showed the highest riskof breakage (75%). Conclusions This study investigated biomechanical stabilityof the novel BLP in relation to properties of the supportingstructures. It was most biomechanically stable when thecolumn and lengthening shaft were composed of Cobalt-Chrome-alloys (type IV).

키워드

FemurFinite Element ModelBone Lengthening Plate

식별자

KCI
ART002182933