A Method to Enhance the Spatial Resolution of the Two-dimensional Detector Arrays for the Precise Dose Assessment of Intensity Modulated Radiation Therapy
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCISCIEScopus
- 저널 순위
- JCR Q4Scopus Q3
- 저널
- JOURNAL OF THE KOREAN PHYSICAL SOCIETY75(12) · pp.1048 · 2019
- 피인용
- Web of Science 1Scopus 1KCI 1
- FWCI (Scopus)
- 0
- JCR IF
- JCR 2018 0.63
- ISSN
- 0374-4884
저자
- 전호상제1저자Department of Radiation Oncology and Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, 50612, Korea
- 주지현Department of Radiation Oncology and Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, 50612, Korea
- 기용간Department of Radiation Oncology and Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Yangsan, 50612, Korea
- 김원택Pusan National University School of Medicine, Yangsan, 50612, Korea
- 남지호Department of Radiation Oncology and Biomedical Research Institute, Pusan National University Hospital, Busan, 49241, Korea
- 김동현Department of Radiation Oncology and Biomedical Research Institute, Pusan National University Hospital, Busan, 49241, Korea
- 박달교신저자Department of Radiation Oncology and Biomedical Research Institute, Pusan National University Hospital, Busan, 49241, Korea
초록
yy Two-dimensional (2D) detector arrays are widely used in the patient-specific quality assurance of intensity modulated radiation therapy (IMRT). One of the disadvantages of the 2D detector arrays is the low spatial resolution. We proposed a new method to enhance the spatial resolution of the 2D detector arrays. We showed that this method also reduced the volume effect of a detector element. To demonstrate the method we evaluated one field of an IMRT verification plan with the 2D ion chamber array. An open field was measured and its penumbra width was evaluated to show the reduction of the volume effect of the detector. Using the proposed method, the distance between measurement points was reduced from 7.62 mm to 6.00 mm and the penumbral width was also reduced.
키워드
2D detector arraySpatial resolutionVolume effectIntensity modulated radiation therapyPatient specific quality assuranceION-CHAMBER ARRAYIMRTVERIFICATIONMATRIXX