Early Detection of Intravenous Infiltration Using Multi-frequency Bioelectrical Impedance Measurement System: Pilot Study
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCIScopus
- 저널
- Journal of Information and Communication Convergence Engineering15(2) · pp.123 · 2017
- 피인용
- Scopus 3KCI 1
- FWCI (Scopus)
- 0
- ISSN
- 2234-8255
저자
초록
The early detection of intravenous (IV) infiltration is necessary to minimize the injury caused by the infiltration, which is one of the most important tasks for nurses. For detecting early infiltration in patients receiving invasive vein treatment, bioelectrical impedance was measured using multi-frequency bioelectrical impedance. The impedance decreased significantly at infiltration, and then decreased gradually over time after infiltration. The relative impedance at 20 kHz decreased remarkably at infiltration, and then gradually decreased thereafter. In addition, the impedance ratio increased temporarily at infiltration and then gradually decreased over time. Furthermore, the impedance at each frequency decreased quantitatively over time. This indicates that IV solution leaking from the vein due to infiltration accumulates in the subcutaneous tissues. Moreover, slopes of log Z vs. log f differently decreased with increasing log f, indicating that the impedance exhibits different responses depending on the frequency.
키워드
Early detectionImpedanceImpedance ratioInfiltrationMulti-frequency impedanceRelative impedance