본문 바로가기

Combined Signal Amplification Using a Propagating Cascade Reaction and a Redox Cycling Reaction for Sensitive Thyroid-Stimulating Hormone Detection

문서 유형
연구출판물 · 학술지 논문
등재
Nature IndexSCIEScopus
저널 순위
JCR Q1Scopus Q1
저널
ANALYTICAL CHEMISTRY91(12) · pp.7894 · 2019
피인용
Web of Science 36Scopus 36
FWCI (Scopus)
1.07
JCR IF
JCR 2018 6.35
ISSN
0003-2700

저자

  1. 박선화제1저자Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan 46241, Korea
  2. 김지현Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan 46241, Korea
  3. 김세헌Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan 46241, Korea
  4. 김경호Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan 46241, Korea
  5. Lee, Nam-SihkEONE Laboratories, Incheon 22014, Korea
  6. Yoon, Young HoEONE Laboratories, Incheon 22014, Korea
  7. 양해식교신저자Department of Chemistry and Chemistry Institute for Functional Materials, Pusan National University, Busan 46241, Korea

초록

Propagating cascade reactions based on two proteases are promising for obtaining high signal amplification. However, in many cases, biosensors that use cascade reactions do not have low detection limits because of the inherent slowness of proteolytic reactions. Here, we report a sensitive electrochemical immunosensor using a high-signal-amplification method that combines a propagating cascade reaction and a redox cycling reaction. The cascade reaction uses ecarin and prothrombin: the ecarin label proteolytically converts inactive prothrombin into active thrombin, which then proteolytically liberates electroactive p-aminophenol (AP) from an AP-conjugated peptide. The liberated AP is electrochemically oxidized to p-benzoquinone imine (QI), regenerated by the reduction of QI by NADH, and then electrochemically reoxidized. This electrochemical-chemical (EC) redox cycling reaction significantly increases the electrochemical signal. The developed immunosensor is also compared with an immunosensor that uses only a propagating cascade reaction and an immunosensor that uses a single proteolytic reaction and an EC redox cycling reaction. The detection limits for thyroid-stimulating hormone (TSH) obtained using the three immunosensors are 3 pg/mL, 2 ng/mL, and 4 ng/mL, respectively, indicating that the newly developed immunosensor is more sensitive than the other two. The measured concentrations of TSH in clinical serum are found to agree well with those determined using a commercial instrument.

키워드

ALKALINE-PHOSPHATASEELECTROCHEMICAL DETECTIONMAGNETIC BEADSASSAYIMMUNOASSAYIMMUNOSENSORPROTHROMBINACTIVATIONTHROMBINBINDING

식별자

Scopus EID
2-s2.0-85067419000
WoS UID