동일한 방사선량에서 모델기반 반복재구성법을 사용한 80 kVp와 120 kVp의 전산화단층 정맥조영술의 영상 품질에 대한 비교
Comparison of Image Qualities of 80 kVp and 120 kVp CT Venography Using Model-Based Iterative Reconstruction at Same Radiation Dose
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCI
- 저널
- JOURNAL OF THE KOREAN SOCIETY OF RADIOLOGY78(4) · pp.235 · 2018
- 피인용
- KCI 0
- ISSN
- 2951-0805
저자
초록
Purpose: To compare image qualities of 80 kVp CT venography (CTV) and 120 kVp CTV by model-based iterative reconstruction (MBIR) at the same radiation dose. Materials and Methods: Sixty-nine patients that underwent CTV using 80 kVp (36 patients, group 1) or 120 kVp (33 patients, group 2) with MBIR at the same radiation dose were enrolled, and objective and subjective image qualities were assessed independently by two radiologists. Results: Mean vascular enhancement and contrast-to-noise ratio were significantly higher in group 1 than in group 2 for inferior vena cavas, femoral veins, and popliteal veins (p < 0.001), and there was significantly lower objective image noise in group 1 (p < 0.001). Subjective analysis revealed image quality was significantly higher in group 1 and image noise was significantly higher in group 2 (p < 0.001). Mean dose-length products was not significantly lower in group 1 (356.1 ± 153.7 mGy cm) than in group 2 (370.1 ± 77.1 mGy cm) (p = 0.635). Conclusion: CTV at 80 kVp with MBIR is a better protocol than CTV at 120 kVp with MBIR at the same radiation dose.
키워드
Radiation Dosage Multidetector Computed Tomography Lower Extremity Veins
식별자
- KCI
- ART002329956