Conformal Hermetic Sealing of Wireless Microelectronic Implantable Chiplets by Multilayered Atomic Layer Deposition (ALD)
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- Nature IndexSCIEScopus
- 저널 순위
- JCR Q1Scopus Q1
- 저널
- ADVANCED FUNCTIONAL MATERIALS29(5) · 2019
- 피인용
- Web of Science 73Scopus 83
- FWCI (Scopus)
- 3.18
- JCR IF
- JCR 2018 15.621
- ISSN
- 1616-301X
저자
- 정준수제1저자Department of Biomedical Engineering, School of Medicine, Pusan National University, Yangsan, South Korea
- Laiwalla, FarahSchool of Engineering, Brown University, Providence, United States
- Lee, JihunSchool of Engineering, Brown University, Providence, United States
- Ritasalo, RiinaPicosun Oy, Espoo, Finland
- Pudas, MarkoPicosun Oy, Espoo, Finland
- Larson, LawrenceSchool of Engineering, Brown University, Providence, United States
- Leung, VincentDepartment of Electrical Engineering, University of California, San Diego, United States
- Nurmikko, Arto교신저자School of Engineering, Brown University, Providence, United States
초록
A hermetic sealing method of sub-millimeter-sized microelectronic chiplets for wireless body implants is presented by ultrathin and electromagnetically transparent atomic layer deposition (ALD) coatings. Fully 3D conformal encapsulation of wirelessly powered microdevices is demonstrated both with and without opening windows for electrophysiological measurements. The chiplets embedding custom application-specific integrated circuits (ASICs) with radio frequency (RF) transmitters are encapsulated by a stack of alternating layers of hafnium oxide and silicon dioxide to maximize impermeability of water and ionic penetration while minimizing the volume of the packaging material. The hermeticity of the devices is characterized through accelerated aging tests in saline at T = 87 degrees C, while continued functionality is monitored via evaluation of backscattered RF signals (near 1 GHz) to ascertain possible degradation and electronic failure. Earliest failures of wirelessly functional devices occur after more than 180 d of immersion at 87 degrees C. Wireless devices having opening windows through the ALD envelope show no signs of degradation for >100 d. This implies an equivalent lifetime >10 years at T = 37 degrees C. This approach is readily scalable to high throughput batch processing of hundreds of microchiplets, offering a methodology for hermetic packaging of microscale biomedical chronic implants.