Linear analysis of single-shot dual-energy computed tomography with a multilayer detector
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q3Scopus Q2
- 저널
- JOURNAL OF INSTRUMENTATION14 · 2019
- 피인용
- Web of Science 4Scopus 4
- FWCI (Scopus)
- 0.54
- JCR IF
- JCR 2018 1.366
- ISSN
- 1748-0221
저자
- 하승우제1저자School of Mechanical Engineering, Pusan National University, Busan, Geumjeong-gu, South Korea
- 김진우School of Mechanical Engineering, Pusan National University, Busan, Geumjeong-gu, South Korea
- 윤종희School of Mechanical Engineering, Pusan National University, Busan, Geumjeong-gu, South Korea
- 김동운School of Mechanical Engineering, Pusan National University, Busan, Geumjeong-gu, South Korea
- 김승호Korea Aerospace Industries, Ltd., Sacheon, Gyeongsangnam-do, South Korea
- 김호경교신저자School of Mechanical Engineering, Pusan National University, Busan, Geumjeong-gu, South Korea
초록
The sandwich-like multilayer detector can have a unique characteristic when it employs two detector layers having different spatial-resolution properties to each other. Reconstruction of dual-energy (DE) images using a weighted log-subtraction technique then results in the unsharp masking-like effect on the resultant images. We have developed a simple model describing the signal and noise in the sandwich detector-based DE computed tomography (CT) using a linear-systems theory. Applying the sandwich detector to a laboratory micro-CT system, we measured the Fourier metrics, such as modulation-transfer function, noise-power spectrum, and noise-equivalent number of quanta (NEQ), using thin wire and water phantoms. The agreements between the measured Fourier metrics and the theoretical models were excellent. The DE-CT for a postmortem mouse showed an enhanced details of bone structures, and the theoretical model explained the reason as showing a shift of the peak position of NEQ to high-frequency regions.
키워드
Computerized Tomography (CT) and Computed Radiography (CR)Image reconstruction in medical imagingOPTIMIZATIONCT