Comparison of Color Light-Emitting Diode Corneal Topographer and Dual Rotating Scheimpflug-Placido Topographer
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q3Scopus Q2
- 저널
- JOURNAL OF OPHTHALMOLOGY2018 · 2018
- 피인용
- Web of Science 3Scopus 6
- FWCI (Scopus)
- 0.73
- JCR IF
- JCR 2017 1.68
- ISSN
- 2090-004X
저자
- Lee, Jae Hyuck제1저자Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, South Korea
- Lee, Yong WooDepartment of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, South Korea
- 이종수Department of Ophthalmology, Pusan National University Hospital, Pusan National University School of Medicine and Medical Research Institute, Yangsan, South Korea
- Knorz, Michael C.Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
- Auffarth, Gerd U.International Vision Correction Research Centre (IVCRC), David J. Apple International Laboratory on Ocular Pathology, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany
- Choi, Chul Young교신저자Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, South Korea
초록
Purpose. To compare measurements of anterior and posterior keratometric values, using a color light-emitting diode corneal topographer and a dual rotating Scheimpflug-Placido topographer. Methods. Anterior and posterior corneal index measurements were performed using a color light-emitting diode corneal topographer (Cassini) and a dual rotating Scheimpflug-Placido topographer (Galilei G4) and then compared. The paired t-test, intraclass correlation coefficients (ICCs), and Bland-Altman plots were used to evaluate the agreement between measurements. Results. Sixty postrefractive surgery eyes and 60 normal eyes were evaluated. Both the color light-emitting diode corneal topographer and the dual rotating Scheimpflug-Placido topographer provided highly repeatable corneal measurements (ICC > 0.969). The agreement levels between the 2 devices for anterior corneal power, astigmatism magnitude, and J0 and J45 values were ICC > 0.906 for the total group. However, the ICC values for posterior corneal power, astigmatism magnitude, and J0 and J45 values were lower than 0.681 for the total group. Conclusions. The anterior keratometric values obtained by the color light-emitting diode corneal topographer and the dual rotating Scheimpflug-Placido topographer showed high agreement levels, but the posterior keratometric values showed lower agreement levels.