Influence of Heart Rate and Innovative Motion-Correction Algorithm on Coronary Artery Image Quality and Measurement Accuracy Using 256-Detector Row Computed Tomography Scanner: Phantom Study
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCISCIEScopus
- 저널 순위
- JCR Q1Scopus Q1
- 저널
- KOREAN JOURNAL OF RADIOLOGY20(1) · pp.94 · 2019
- 피인용
- Web of Science 8Scopus 8KCI 2
- FWCI (Scopus)
- 0.4
- JCR IF
- JCR 2018 3.73
- ISSN
- 1229-6929
저자
- 박정빈제1저자Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
- 정연주Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
- 이기원Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
- 이남경Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
- 김진유Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
- 이지원A교신저자Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
초록
Objective: To investigate the efficacy of motion-correction algorithm (MCA) in improving coronary artery image quality and measurement accuracy using an anthropomorphic dynamic heart phantom and 256-detector row computed tomography (CT) scanner. Materials and Methods: An anthropomorphic dynamic heart phantom was scanned under a static condition and under heart rate (HR) simulation of 50-120 beats per minute (bpm), and the obtained images were reconstructed using conventional algorithm (CA) and MCA. We compared the subjective image quality of coronary arteries using a four-point scale (1, excellent; 2, good; 3, fair; 4, poor) and measurement accuracy using measurement errors of the minimal luminal diameter (MLD) and minimal luminal area (MLA). Results: Compared with CA, MCA significantly improved the subjective image quality at HRs of 110 bpm (1.3 +/- 0.3 vs. 1.9 +/- 0.8, p = 0.003) and 120 bpm (1.7 +/- 0.7 vs. 2.3 +/- 0.6, p = 0.006). The measurement error of MLD significantly decreased on using MCA at 110 bpm (11.7 +/- 5.9% vs. 18.4 +/- 9.4%, p = 0.013) and 120 bpm (10.0 +/- 7.3% vs. 25.0 +/- 16.5%, p = 0.013). The measurement error of the MLA was also reduced using MCA at 110 bpm (19.2 +/- 28.1% vs. 26.4 +/- 21.6%, p = 0.028) and 120 bpm (17.9 +/- 17.7% vs. 34.8 +/- 19.6%, p = 0.018). Conclusion: Motion-correction algorithm can improve the coronary artery image quality and measurement accuracy at a high HR using an anthropomorphic dynamic heart phantom and 256-detector row CT scanner.