Influence of Heart Rate and Innovative Motion-Correction Algorithm on Coronary Artery Image Quality and Measurement Accuracy Using 256-Detector Row Computed Tomography Scanner: Phantom Study

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q1Scopus Q1
저널
KOREAN JOURNAL OF RADIOLOGY20(1) · pp.94 · 2019
피인용
Web of Science 8Scopus 8KCI 2
FWCI (Scopus)
0.4
JCR IF
JCR 2018 3.73
ISSN
1229-6929

저자

  1. 박정빈제1저자Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
  2. 정연주Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
  3. 이기원Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
  4. 이남경Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
  5. 김진유Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea
  6. 이지원A교신저자Department of Radiology, Pusan National University School of Medicine and Medical Research Institute, Pusan National University Hospital, Busan, South Korea

초록

Objective: To investigate the efficacy of motion-correction algorithm (MCA) in improving coronary artery image quality and measurement accuracy using an anthropomorphic dynamic heart phantom and 256-detector row computed tomography (CT) scanner. Materials and Methods: An anthropomorphic dynamic heart phantom was scanned under a static condition and under heart rate (HR) simulation of 50-120 beats per minute (bpm), and the obtained images were reconstructed using conventional algorithm (CA) and MCA. We compared the subjective image quality of coronary arteries using a four-point scale (1, excellent; 2, good; 3, fair; 4, poor) and measurement accuracy using measurement errors of the minimal luminal diameter (MLD) and minimal luminal area (MLA). Results: Compared with CA, MCA significantly improved the subjective image quality at HRs of 110 bpm (1.3 +/- 0.3 vs. 1.9 +/- 0.8, p = 0.003) and 120 bpm (1.7 +/- 0.7 vs. 2.3 +/- 0.6, p = 0.006). The measurement error of MLD significantly decreased on using MCA at 110 bpm (11.7 +/- 5.9% vs. 18.4 +/- 9.4%, p = 0.013) and 120 bpm (10.0 +/- 7.3% vs. 25.0 +/- 16.5%, p = 0.013). The measurement error of the MLA was also reduced using MCA at 110 bpm (19.2 +/- 28.1% vs. 26.4 +/- 21.6%, p = 0.028) and 120 bpm (17.9 +/- 17.7% vs. 34.8 +/- 19.6%, p = 0.018). Conclusion: Motion-correction algorithm can improve the coronary artery image quality and measurement accuracy at a high HR using an anthropomorphic dynamic heart phantom and 256-detector row CT scanner.