Mass measurement method based on resonance frequency of the capacitive thin membrane sensor

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q3Scopus Q2
저널
JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY32(7) · pp.3263 · 2018
피인용
Web of Science 4Scopus 5KCI 3
FWCI (Scopus)
0.36
JCR IF
JCR 2017 1.194
ISSN
1738-494X

저자

  1. 박영주제1저자Electronic Engineering Department, Pusan National University, Busan, South Korea
  2. 김상만Electronic Engineering Department, Pusan National University, Busan, South Korea
  3. 박주성교신저자Electronic Engineering Department, Pusan National University, Busan, South Korea

초록

This paper introduces a mass measurement method and a system for capacitive membrane sensor, which uses the principle of membrane displacement at a resonance frequency which maximizes output voltage due to its strong vibration. The implemented system consists of dedicated hardware, firmware, and PC. The resonance frequencies of the Si3N4 membrane film are 59.2 kHz and 58.2 kHz measured using laser vibrometer and the proposed method, respectively. The resonance frequency of which quartz has a resonance frequency of 32.768 kHz is 32.775 kHz by our method. The suggested method has 600 times more sensitive in measuring the resonance frequency of a capacitive membrane than the conventional static capacitance measurement method.

키워드

Capacitive membrane sensorResonance frequencyMass measurementCMUTCHEMICAL SENSORMICRO-MEMBRANEOSCILLATORBIOSENSORSOPTIMIZATIONTRANSDUCERSYSTEMARRAY