본문 바로가기

Development of an Attachable Endoscopic Nerve Stimulator for Intraoperative Neuromonitoring during Endoscopic or Robotic Thyroidectomy

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q1
저널
OTOLARYNGOLOGY-HEAD AND NECK SURGERY158(3) · pp.465 · 2018
피인용
Web of Science 11Scopus 13
FWCI (Scopus)
1.58
JCR IF
JCR 2017 2.444
ISSN
0194-5998

저자

  1. 성의숙제1저자Department of Otolaryngology–Head and Neck Surgery and Research Institute for Convergence of Biomedical Science and Technology, College of Medicine, Pusan National University, Yangsan, South Korea
  2. 이진춘Department of Otolaryngology–Head and Neck Surgery and Research Institute for Convergence of Biomedical Science and Technology, College of Medicine, Pusan National University, Yangsan, South Korea
  3. 김석현Department of Otolaryngology–Head and Neck Surgery and Research Institute for Convergence of Biomedical Science and Technology, College of Medicine, Pusan National University, Yangsan, South Korea
  4. 신성찬Department of Otolaryngology–Head and Neck Surgery, College of Medicine and Medical Research Institute, Pusan National University, Pusan, South Korea
  5. 정다운Department of Otolaryngology–Head and Neck Surgery, College of Medicine and Medical Research Institute, Pusan National University, Pusan, South Korea
  6. 이병주교신저자Department of Otolaryngology–Head and Neck Surgery, College of Medicine and Medical Research Institute, Pusan National University, Pusan, South Korea

초록

We developed a simple attachable endoscopic nerve stimulator that can be connected to monopolar cauterization surgical instruments. This study on porcine models aimed to investigate the feasibility and efficacy of an attachable endoscopic nerve stimulator for intraoperative neuromonitoring (IONM) before application in humans. We evaluated the electromyography (EMG) amplitudes of 8 recurrent laryngeal nerves in 4 pigs with a conventional nerve probe and the attachable endoscopic nerve stimulator. The attachable endoscopic nerve stimulator was feasible and safe in all cases. There was no significant difference in the EMG amplitude of the recurrent laryngeal nerve among instruments (P = .429). The application of stimulating dissection with an attachable endoscopic nerve stimulator during endoscopic or robotic thyroidectomy with IONM is simple, convenient, and effective. It provides surgeons with real-time feedback of the EMG response during intermittent IONM. We believe that this novel device could be an essential guide and functional navigator for most surgeons, especially for less experienced ones.

키워드

attachable nerve stimulatorendoscopyrobotneuromonitoringrecurrent laryngeal nervethyroidectomySURGERY

식별자

Scopus EID
2-s2.0-85042880882
WoS UID