Image quality and radiation dose of CT venography with double dose reduction using model based iterative reconstruction: comparison with conventional CT venography using filtered back projection
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q3Scopus Q2
- 저널
- ACTA RADIOLOGICA59(5) · pp.546 · 2018
- 피인용
- Web of Science 4Scopus 4
- FWCI (Scopus)
- 0.2
- JCR IF
- JCR 2017 1.823
- ISSN
- 0284-1851
저자
- 정여진제1저자Department of Radiology, Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Hospital, Busan, South Korea
- 추기석교신저자Department of Radiology, Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Hospital, Busan, South Korea
- 남경진Department of Radiology, Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Hospital, Busan, South Korea
- 이지원ADepartment of Radiology, Pusan National University Hospital, Pusan National University School of Medicine, Busan, South Korea
- 김진유Department of Radiology, Pusan National University Hospital, Pusan National University School of Medicine, Busan, South Korea
- 정혁재Department of Vascular surgery, Pusan National University Yangsan Hospital, Pusan National University School of Medicine, Busan, South Korea
- Lim, Soo JinDepartment of Cardiology, Kim Hae Gang-il Hospital, KyoungNam, South Korea
초록
Background: Computed tomography venography (CTV) at low kVp using model-based iterative reconstruction (MBIR) can enhance vascular enhancement with noise reduction. Purpose: To evaluate image qualities and radiation doses of CTV at 80 kVp using MBIR and a small iodine contrast media (CM) dose and to compare these with those of CTV performed using a conventional protocol. Material and Methods: Sixty-five patients (mean age = 58.1 +/- 7.2 years) that underwent CTV for the evaluation of deep vein thrombosis (DVT) and varicose veins were enrolled in this study. Patients were divided into two groups: Group A (35 patients, 80 kVp, MBIR, automatic tube current modulation, CM = 270 mg/mL, 100 mL) and Group B (30 patients, 100 kVp, filtered back projection [FBP], 120 fixed mA, CM = 370 mg/mL, 120 mL). Objective and subjective image qualities of inferior vena cava (IVC), femoral vein (FV), and popliteal vein (PV) were assessed and radiation doses were recorded. Results: Mean vascular enhancement in group A was significantly lower than in group B (P < 0.01). Noise in group A was significantly lower than in group B except for PV and contrast-to-noise ratio were not significantly different in the two groups (P > 0.05). In addition, radiation dose in group A was significantly lower than in group B (P < 0.001). Subjective image quality comparison revealed group A was statistically inferior to group B except for subjective image noise. Conclusion: CTV at 80 kVp using MBIR with small iodine contrast dose provided acceptable image quality at a lower radiation dose than conventional CTV using FBP.
키워드
Diagnosisradiographymultidetector computed tomographylower extremityveinsLOW KILOVOLTAGE SETTINGSDEEP VENOUS THROMBOSISCOMPUTED-TOMOGRAPHYPULMONARY ANGIOGRAPHYVASCULAR ENHANCEMENTMULTIDETECTOR CTMODULATIONSONOGRAPHYSTANDARDKVP