Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCISCIEScopus
- 저널 순위
- JCR Q4Scopus Q3
- 저널
- JOURNAL OF THE KOREAN PHYSICAL SOCIETY66(12) · pp.1926 · 2015
- 피인용
- Web of Science 13KCI 1
- JCR IF
- JCR 2014 0.418
- ISSN
- 0374-4884
저자
초록
We investigated the performances of two digital mammography systems. The systems use a cesium-iodide (CsI) scintillator and an amorphous selenium (a-Se) photoconductor for X-ray detection and are installed in the same hospital. As physical metrics, we measured the modulationtransfer function (MTF), the noise-power spectrum (NPS), and the detective quantum efficiency (DQE). In addition, we analyzed the contrast-detail performances of the two systems by using a commercial contrast-detail phantom. The CsI-based indirect conversion detector provided better MTF and DQE performances than the a-Se-based direct conversion detector whereas the former provided a poorer NPS performance than the latter. These results are explained by the fact that the CsI-based detector used an MTF restoration preprocessing algorithm. The a-Se-based detector showed better contrast-detail performance than the CsI-based detector. We believe that the highfrequency noise characteristic of a detector is more responsible for the visibility of small details than its spatial-resolution performance.
키워드
Digital mammographyCesium iodideAmorphous seleniumMTFNPSDQEContrastMODULATION TRANSFER-FUNCTIONPHYSICAL-CHARACTERISTICSIMAGING PERFORMANCEBREAST TOMOSYNTHESISFILM MAMMOGRAPHYRAYSELENIUMSIGNALRESOLUTIONPROTOTYPE