Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q4Scopus Q3
저널
JOURNAL OF THE KOREAN PHYSICAL SOCIETY66(12) · pp.1926 · 2015
피인용
Web of Science 13KCI 1
JCR IF
JCR 2014 0.418
ISSN
0374-4884

저자

  1. 윤한빈제1저자부산대학교, 공과대학, 기계공학부, KO
  2. 한종철부산대학교, 공과대학, 기계공학부, KO
  3. 윤승만부산대학교, 공과대학, 기계공학부, KO
  4. 감수화부산대학교, 공과대학, 기계공학부, KO
  5. Cho, SeungryongKAIST, Dept Nucl & Quantum Engn, KO
  6. 김호경교신저자부산대학교, 공과대학, 기계공학부, 제어자동화시스템전공, KO

초록

We investigated the performances of two digital mammography systems. The systems use a cesium-iodide (CsI) scintillator and an amorphous selenium (a-Se) photoconductor for X-ray detection and are installed in the same hospital. As physical metrics, we measured the modulationtransfer function (MTF), the noise-power spectrum (NPS), and the detective quantum efficiency (DQE). In addition, we analyzed the contrast-detail performances of the two systems by using a commercial contrast-detail phantom. The CsI-based indirect conversion detector provided better MTF and DQE performances than the a-Se-based direct conversion detector whereas the former provided a poorer NPS performance than the latter. These results are explained by the fact that the CsI-based detector used an MTF restoration preprocessing algorithm. The a-Se-based detector showed better contrast-detail performance than the CsI-based detector. We believe that the highfrequency noise characteristic of a detector is more responsible for the visibility of small details than its spatial-resolution performance.

키워드

Digital mammographyCesium iodideAmorphous seleniumMTFNPSDQEContrastMODULATION TRANSFER-FUNCTIONPHYSICAL-CHARACTERISTICSIMAGING PERFORMANCEBREAST TOMOSYNTHESISFILM MAMMOGRAPHYRAYSELENIUMSIGNALRESOLUTIONPROTOTYPE

식별자

WoS UID
000357675100023
KCI
ART002002581