The Effect of Corneal Biomechanical Properties on Rebound Tonometer in Patients With Normal-Tension Glaucoma

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q1Scopus Q1
저널
AMERICAN JOURNAL OF OPHTHALMOLOGY159(1) · pp.144 · 2015
피인용
Web of Science 39
JCR IF
JCR 2014 3.871
ISSN
0002-9394

저자

  1. 신종훈제1저자Department of Ophthalmology, Pusan National University, School of Medicine, Seo-gu-Busan, South Korea
  2. 이지웅교신저자Department of Ophthalmology, Pusan National University, School of Medicine, Seo-gu-Busan, South Korea
  3. 김은아Jules Stein Eye Institute, University of California at Los Angeles, School of Medicine, Los Angeles, United States
  4. Caprioli, JosephJules Stein Eye Institute, University of California at Los Angeles, School of Medicine, Los Angeles, United States

초록

center dot PURPOSE: To evaluate the effects of corneal biomechanical properties on intraocular pressure (IOP) measured with the ICare, and to compare IOP readings obtained with ICare, Ocular-Response Analyzer (ORA), and Goldmann applanation tonometry (GAT) in normal-tension glaucoma (NTG) and normal subjects. center dot DESIGN: Prospective, cross-sectional, comparative study. center dot METHODS: IOP was measured with ICare, ORA, and GAT. All subjects had corneal hysteresis (CH) and corneal resistance factor (CRF), which were measured with ORA; and central corneal thickness (CCT), axial length, spherical equivalent, and keratometry. center dot RESULTS: This study enrolled 97 eyes of 97 NTG patients and 89 eyes of 89 normal subjects. CCT, CH, and CRF in NTG patients were significantly lower than those in normal subjects (P = .033, P = .006, and P = .003). The difference in IOP between techniques was highly significant in NTG patients (P < .001), while there was no significant difference in IOP values between techniques in normal controls (P = .931). ICare readings were significantly lower than corneal-compensated IOP in NTG patients (P = .014). CH and CRF were significantly associated with IOP measurements with ICare in NTG and normal subjects (P < .001). The greater difference between IOPcc and ICare in NTG patients was significantly influenced by the lower CH (P < .001). center dot CONCLUSIONS: Since ICare is a convenient way to measure IOP, ICare is a reasonable option as an alternative tonometer in NTG patients. However, the clinician must consider that the corneal biomechanical characteristics in NTG can cause ICare to underestimate IOP. (C) 2015 by Elsevier Inc. All rights reserved.

키워드

OCULAR-RESPONSE-ANALYZEROPEN-ANGLE GLAUCOMAGOLDMANN APPLANATION TONOMETERINTRAOCULAR-PRESSURE MEASUREMENTSNORMAL EYESTHICKNESSHYSTERESISLATANOPROSTHYPERTENSIONPROGRESSION