본문 바로가기

Bio-signal acquisition circuit with high signal-to-noise ratio for u-healthcare system

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q3Scopus Q2
저널
ELECTRONICS LETTERS50(23) · pp.1671 · 2014
피인용
Web of Science 1
JCR IF
JCR 2013 1.068
ISSN
0013-5194

저자

  1. 남일구제1저자Department of Electronic Engineering, PNU-LG Electronics Smart Control Center, Pusan National University, Busan, South Korea
  2. Kang, HongseokDepartment of ICEE, Catholic University of Korea, Bucheon-si, Gyeonggi-do, South Korea
  3. Woo, Doo HyungDepartment of ICEE, Catholic University of Korea, Bucheon-si, Gyeonggi-do, South Korea

초록

The testing of a monolithic bio-signal acquisition circuit with an analogue negative feedback loop for ubiquitous healthcare (u-healthcare) and the monitoring of physical activity is described. The large low-frequency signal is effectively rejected by the proposed circuit, allowing the dynamic range of the more valuable signals to be improved. The proposed circuit has been designed and fabricated using a 0.18 mu m 1-poly 6-metal CMOS process. Using an analogue negative feedback loop with a dual-gain filter, the proposed method provides an average signal-to-noise ratio 8.1 dB higher than that of the conventional method.