본문 바로가기

Thick-lens velocity-map imaging spectrometer with high resolution for high-energy charged particles

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2Scopus Q2
피인용
Web of Science 73
JCR IF
JCR 2013 1.526
ISSN
1748-0221

저자

  1. Kling, N. G.제1저자J.R. Macdonald Laboratory, Department of Physics, United States
  2. Paul, D.Physics Department, CASTECH, POSTECH, South Korea
  3. Gura, A.J.R. Macdonald Laboratory, Department of Physics, United States
  4. Laurent, G.J.R. Macdonald Laboratory, Department of Physics, United States
  5. De, S.J.R. Macdonald Laboratory, Department of Physics, United States
  6. Li, H.J.R. Macdonald Laboratory, Department of Physics, United States
  7. Wang, Z.J.R. Macdonald Laboratory, Department of Physics, United States
  8. Ahn, B.Physics Department, CASTECH, POSTECH, South Korea
  9. Kim, C. H.Physics Department, CASTECH, POSTECH, South Korea
  10. 김태규Department of Chemistry, Chemistry Institute of Functional Materials, Pusan National University, South Korea
  11. Litvinyuk, I. V.J.R. Macdonald Laboratory, Department of Physics, United States
  12. Cocke, C. L.J.R. Macdonald Laboratory, Department of Physics, United States
  13. Ben-Itzhak, I.J.R. Macdonald Laboratory, Department of Physics, United States
  14. Kim, D.Physics Department, CASTECH, POSTECH, South Korea
  15. Kling, M. F.교신저자J.R. Macdonald Laboratory, Department of Physics, United States

초록

A novel design for a velocity-map imaging (VMI) spectrometer with high resolution over a wide energy range surpassing a standard VMI design is reported. The main difference to a standard three-electrode VMI is the spatial extension of the applied field using 11 electrodes forming a thick-lens. This permits measurements of charged particles with higher energies while achieving excellent resolving power over a wide range of energies. Using SIMION simulations, the thick-lens VMI is compared to a standard design for up to 360 eV electrons. The simulations also show that the new spectrometer design is suited for charged-particle detection with up to 1 keV using a repeller-electrode voltage of -30 kV. The experimental performance is tested by laser-induced ionization of rare gases producing electrons up to about 70 eV. The thick-lens VMI is useful for a wide variety of studies on atoms, molecules and nanoparticles in intense laser fields and high-photon-energy fields from high-harmonic-generation or free-electron lasers.

키워드

SpectrometersImaging spectroscopyInstrumentation for FELABOVE-THRESHOLD IONIZATIONMOMENTUM SPECTROSCOPYRECOIL-IONDYNAMICSINVERSIONXENON

식별자

WoS UID
000340036100075