본문 바로가기

Percutaneous Retrieval and Redeployment of an Atrial Septal Occluder under Three-Dimensional Transesophageal Echocardiographic Guidance: A Case Report

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q3Scopus Q2
저널
JOURNAL OF KOREAN MEDICAL SCIENCE29(6) · pp.871 · 2014
피인용
Web of Science 4KCI 0
JCR IF
JCR 2013 1.253
ISSN
1011-8934

저자

  1. 안진희부산대학교
  2. 김진희부산대학교
  3. 최정현부산대학교
  4. 오준혁교신저자부산대학교

초록

Percutaneous device closure for secundum atrial septal defects (ASDs) has been performed commonly and safely with high success rates. However, it is still challenging to close ASDs that are surrounded with deficient or hypermobile rims and could be compromised with an unexpected migration of device. We report a case of percutaneous Amplazter Septal Occluder (ASO; St. Jude Medical Inc., St. Paul, Minnesota, USA) device closure for an ASD with a thin and floppy interatrial septum, which immediately migrated into the right atrium and was not pulled back into the delivery sheath. To our knowledge, this is the first report on a successful percutaneous retrieval and redeployment of the device in such a situation, preventing any vascular injury or unplanned emergency open heart surgery.

키워드

Atrial Septal OccluderMigrationRetrievalPercutaneousTRANSCATHETERCLOSUREDEVICEADULTSDEFECT

식별자

WoS UID
000337339300021
KCI
ART001881515