Pressure-wire based assessment of microvascular resistance using calibrated upstream balloon obstruction
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q2Scopus Q1
- 저널
- CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS80(4) · pp.581 · 2012
- 피인용
- Web of Science 8
- JCR IF
- JCR 2011 2.29
- ISSN
- 1522-1946
저자
- 김준홍제1저자부산대학교, 의과대학, 의학과, KO
- 박주현부산대학교, 부산대학교병원, 대학병원, KO
- 추기석부산대학교, 의과대학, 의학과, KO
- 송성국부산대학교, 의과대학, 의학과, KO
- 김정수부산대학교, 의과대학, 의학과, KO
- 박용현부산대학교, 의과대학, 의학과, KO
- 김준부산대학교, 의과대학, 의학과, KO
- 전국진부산대학교, 의과대학, 의학과, 내과학교실, KO
- 한동철부산대학교, 부산대학교병원, 대학병원, KO
- Faranesh, Anthony Z.
- Lederman, Robert J.교신저자NHLBI, Cardiovasc & Pulm Branch, Div Intramural Res, US
초록
Objectives: We assess microvascular integrity as a marker of myocardial viability after coronary stenting, using only a pressure guidewire. Background: Microvascular integrity generally is not assessed using pressure-only guidewires because the transducer lies upstream of microvasculature. We partially inflate a balloon inside a coronary stent to achieve a specific normalized pressure drop at rest (distal coronary/aortic pressure = 0.8) and then infuse a vasodilator, to render the wire sensitive to microvascular function. We hypothesize that the further decline in pressure (?FFR0.8) predicts MRI myocardial viability. Methods: We studied 29 subjects with acute coronary syndrome including myocardial infarction. After successful culprit stenting, the resting coronary/aortic pressure was set to 0.8 using temporary balloon obstruction. ?FFR0.8 was defined as 0.8-(distal coronary/aortic pressures) during adenosine-induced hyperemia. The average transmural extent of infarction was defined as the average area of MRI late gadolinium enhancement (after 2.8 +/- 1.5 days) divided by the corresponding full thickness of the gadolinium enhanced sector in short axis slices, and was compared with ?FFR0.8. Results: ?FFR0.8 corresponded inversely and linearly with the average transmural extent of infarction (r2 = 0.65, P < 0.001). We found that a transmural extent of infarction of 0.50 corresponded to a ?FFR0.8 threshold of 0.1, and had high sensitivity and specificity (100% and 94.4%, respectively). Conclusions: Using only an upstream pressure-sensitive guidewire and a partially obstructing balloon during pharmacologic hyperemia, we were able to predict MRI myocardial viability with high accuracy after relief of epicardial stenosis. With further validation, this may prove a useful clinical prognostic tool after percutaneous intervention. (c) 2011 Wiley Periodicals, Inc.
키워드
coronary artery physiologymicrovascular resistancemagnetic resonance imaginginterventional cardiologyintravascular diagnosticsFRACTIONAL FLOW RESERVEMYOCARDIAL-INFARCTIONMICROCIRCULATORY RESISTANCECORONARY INTERVENTIONVIABLE MYOCARDIUMSTENOSIS SEVERITYCATHETERIZATIONINDEXANGIOPLASTYGUIDELINES
식별자
- WoS UID
- 000308929600016



로그인
연구물 목록으로