Dissolved oxygen concentration field measurement in micro-scale water flows using PtOEP/PS film sensor
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- SCIEScopus
- 저널 순위
- JCR Q2Scopus Q1
- 저널
- OPTICS AND LASERS IN ENGINEERING50(1) · pp.74 · 2012
- 피인용
- Web of Science 29
- JCR IF
- JCR 2011 1.838
- ISSN
- 0143-8166
저자
초록
A planar optode system based on an oxygen quenchable luminophore platinum (II) octaethyporphrin (PtOEP) bound with thin polystyrene (PS) film and UV light-emitting diodes (UV-LEDs) was developed to measure the dissolved oxygen (DO) concentration field in micro-scale water flows. An intensity-based method adopting a pixel-to-pixel in situ calibration technique was used to visualize DO concentration fields around an impinging micro-nozzle. The achievable spatial resolution of the acquired concentration map could be as high as 2.94 mu m. A micro-round water jet having 100% of DO was obliquely impinged on to a PtOEP/PS film coated plate placed in a 0% of DO water container. Velocity fields were obtained by computational fluid dynamics (CFD) analysis and it is demonstrated that the high DO concentration region was coincided with the impingement area. The DO concentration gradient due to DO diffusion was affected by the Reynolds number. (C) 2011 Elsevier Ltd. All rights reserved.
키워드
Dissolved oxygen concentration fieldMicro-jetPtOEP/PS filmOptical sensorPRESSURE-SENSITIVE PAINTSENSING MATERIALPORPHYRIN DYESPLATINUM