Evaluation of an acid-fast bacilli autostainer for concentrated sputum smears

문서 유형
연구출판물 · 학술지 논문
등재
SCIEScopus
저널 순위
JCR Q2
피인용
Web of Science 3
JCR IF
JCR 2007 2.24
ISSN
1027-3719

저자

  1. 김신영제1저자부산대학교, 의과대학, 의학과, KO
  2. 김영진부산대학교, 의과대학, 의학과, KO
  3. 황상현부산대학교, 의과대학, 의학과, 진단검사의학교실, KO
  4. 김형회부산대학교, 의과대학, 의과학과, 의료정보학, KO
  5. 이은엽부산대학교, 의과대학, 의학과, KO
  6. 장철훈교신저자부산대학교, 의과대학, 의학과, KO

초록

SETTING: An automatic acid-fast bacilli (AFB) stainer was developed to improve laboratory workflow and protect laboratory workers. OBJECTIVE: To determine the usefulness of the 7720 Aerospray AFB slide stainer for liquefied and concentrated respiratory specimens. DESIGN: With 24 smear-positive specimens submitted for AFB stain and culture in a tertiary care hospital, the autostainer was evaluated to determine whether smears would be sloughed from slides during the staining process by loading distilled water instead of staining reagents. With 109 more specimens, the performance of the 7720 Aerospray was compared with manual florescence staining as a gold standard. The tested smears were prepared with Mycohold treatment. RESULTS: In most cases, the slides exposed to the autostainer containing distilled water showed lower grades of AFB than those slides not exposed to the autostainer (14/24 in Mycohold-treated slides). Almost half (54/ 109, 49.5 %) of the Mycohold-treated autostained smears had a lower grade than the same smears stained manually. Among 62 Mycohold-treated autostained smear-negative specimens, respectively eight (12.9%) and 19 (30.6%) were high and low false-negatives. CONCLUSION: The 7720 Aerospray showed poor performance in AFB staining of concentrated sputum specimens.

식별자

WoS UID
000251868600009