본문 바로가기

A comparative study of particle size analysis in fine powder: The effect of a polycomponent particulate system

문서 유형
연구출판물 · 학술지 논문
등재
KCISCIEScopus
저널 순위
JCR Q3
저널
KOREAN JOURNAL OF CHEMICAL ENGINEERING26(1) · pp.300 · 2009
피인용
Web of Science 15KCI 8
JCR IF
JCR 2008 0.83
ISSN
0256-1115

저자

  1. 최희규제1저자Changwon National University
  2. Woong LeeChangwon National University
  3. 김동욱Changwon National University
  4. 샤런드라쿠마르창원대학교
  5. 하종학Pusan National University
  6. Seongsoo KimGyeongsang National University
  7. 이정은Pusan National University

초록

We report the particle size distribution of poly component particulate systems studied in three kinds of experimental methods. Six analyzers, such as Mastersizer Microplus (Malvern Instruments Ltd., UK), LS230 (Coulter Electronics Ltd., USA) LMS30 (Seishin, Japan), Analysette22 (Fritsch, Gennany), HELOS (Sympatec, Germany) based on a laser diffraction and scattering method, and the SKC-2000S (Seishin Co., Ltd., Japan) based on the centrifugal sedimentation method, were used to study the particle size distribution. The results do not show reasonably good agreement between the different analyzers and different sample systems. There is a discrepancy regarding the absolute values, which can be explained by the fact the techniques used are based on different measuring principles. The results of the present study reflect that the investigator must carefully select the particle size analyzer for a particular application. Therefore, we suggest that it is necessary to measure the particle size distribution by using at least two types of analyzers or more of different makers.

식별자

WoS UID
000262722500052
KCI
ART001313165