Chromium(III)-selective Electrode Using p-(4-Acetanilidazo)calix[4]arene as an Ionophore in PVC Matrix
- 문서 유형
- 연구출판물 · 학술지 논문
- 등재
- KCISCIEScopus
- 저널 순위
- JCR Q2
- 저널
- Bulletin of the Korean Chemical Society29(12) · pp.2471 · 2008
- 피인용
- Web of Science 27KCI 0
- JCR IF
- JCR 2007 1.156
- ISSN
- 0253-2964
저자
- 심윤보교신저자부산대학교
- Pankaj KumarDepartment of Chemistry, Hindustan College of Science and Technology
초록
A new poly(vinyl chloride) membrane, containing synthesized p-(4-acetanilidazo)calix[4]arene as a membrane carrier, was used to fabricate a chromium(M) selective electrode. The optimum composition of the best performing membrane contained p-(4-acetanilidazo)calix[4]arene:DBBP:NaTPB:PVC in the ratio 10:100:1:100 w/w. This electrode exhibited a near Nernstian slope of 19.8 +/- 0.2 mV/decade of activity in the working concentration range of 9.8 x 10(-7)-1.0 x 10(-1) M. It displayed a stable potential response in tile pH range 2.8-5.7. The electrode exhibited a fast response time of less than 14 s and could be used for at least 12 weeks without any considerable divergence in potentials. This electrode showed very good selectivity over most of the common cations including Li+, K+, Na+, Ni2+, Co2+, Cu2+, Sr2+, Ba2+, Cs+, Pb-,(2+) Zn2+, Mg2+, Cd2+, Al3+, Fe3+, and La3+. It could be employed successfully for the determination of Cr(III) ion activity in real samples.