MONTE CARLO DOSE ASSESSMENT IN DENTAL CONE-BEAM COMPUTED TOMOGRAPHY

Document Type
Publications · Journal Article
Registered DB
SCIEScopus
Journal Ranking
JCR Q4Scopus Q3
Journal
RADIATION PROTECTION DOSIMETRY193(3-4) · pp.190 · 2021
Citation
Web of Science 3Scopus 4
FWCI (Scopus)
0.22
JCR IF
JCR 2020 0.972
ISSN
0144-8420

Authors

  1. Jinwoo KimFirst authorPusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea
  2. Jeon, HosangPusan Natl Univ, Dept Radiat Oncol, Yangsan Hosp, Yangsan Si 50612, Gyeongsangnam D, South Korea
  3. Kim, Ho KyungCorresponding authorPusan Natl Univ, Sch Mech Engn, Busan 46241, South Korea

Abstract

Most dental cone-beam computed tomography (CBCT) uses an x-ray beam field covering the maxillomandibular region and the width-truncated detector geometry. The spatial dose distribution in dental CBCT is analyzed in terms of local primary and remote secondary doses by using a list-mode analysis of x-ray interactions obtained from the Monte Carlo simulations. The patient-dose benefit due to the width-truncated detector geometry is also investigated for a wide range of detector offsets. The developed dose estimation agrees with the measurement in a relative error of 7.7%. The secondary dose outside of the irradiation field becomes larger with increasing tube voltage. The dose benefit with the width-truncated geometry linearly increases as the detector-offset width is decreased. Leaving the CT image quality out of the account, the MC results reveal that the operation of dental CBCT with a lower tube voltage and a smaller detector-offset width is beneficial to the patient dose.

Keywords

TUBE CURRENT MODULATIONIMAGE QUALITYCHEST CTCBCT DEVICESREDUCTIONDOSIMETRYRECONSTRUCTIONOPTIMIZATIONSCANNERSPHANTOM