Skip to content

Quantitative Analysis of Helical Flow with Accuracy Using Ultrasound Speckle Image Velocimetry: In Vitro and in Vivo Feasibility Studies

QUANTITATIVE ANALYSIS OF HELICAL FLOW WITH ACCURACY USING ULTRASOUND SPECKLE IMAGE VELOCIMETRY: &ITIN VITRO&IT AND &ITIN VIVO&IT FEASIBILITY STUDIES

Document Type
Publications · Journal Article
Registered DB
SCIEScopus
Journal Ranking
JCR Q1Scopus Q1
Journal
ULTRASOUND IN MEDICINE AND BIOLOGY44(3) · pp.657 · 2018
Citation
Web of Science 8Scopus 9
FWCI (Scopus)
0.73
JCR IF
JCR 2017 2.645
ISSN
0301-5629

Authors

  1. Lee, Sang JoonFirst authorCorresponding authorDepartment of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, South Korea
  2. Park, Jun HongDepartment of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, South Korea
  3. Kim, Jeong JuDepartment of Mechanical Engineering, Pohang University of Science and Technology (POSTECH), Pohang, South Korea
  4. Yeom, EunseopSchool of Mechanical Engineering, Pusan National University, Busan, South Korea

Abstract

Venous valve dysfunction and induced secondary abnormal flows are closely associated with venous diseases. Thus, detailed analysis of venous valvular flow is invaluable from biological and medical perspectives. However, most of the previous studies on venous perivalvular flows were based on qualitative analysis. On the contrary, quantitative analysis of perivalvular flows has not been fully understood. In this study, we used the ultrasound speckle image velocimetry (SIV) technique, which utilizes the speckle patterns of red blood cells (RBCs) created by ultrasound waves to measure 3-D valvular flows quantitatively. The flow structures obtained with the proposed SIV technique for an in vitro model were compared with those obtained by numerical simulation and the color Doppler method to validate the measurement accuracy of the ultrasound SIV technique. Blood flow in the human great saphenous vein was then measured at various distances from the valve with and without exercise. 3-D valvular flow was analyzed in accordance with the dimensionless index, helical intensity. The results obtained by the proposed method matched well with those obtained by numerical simulation and the color Doppler method. The hemodynamic characteristics of 3-D valvular helical flow which were analyzed experimentally using the SIV method would be used for quantitative diagnosis of venous valvular diseases. (C) 2018 World Federation for Ultrasound in Medicine & Biology. All rights reserved.

Keywords

Ultrasound speckle image velocimetryHelical flowVenous valveHelical intensityBLOOD-FLOW3-DIMENSIONAL FLOWVEINSPATTERNSTRESSSYSTEMBYPASSVALVESPIV